Multi-Point Line Focused Laser Differential Interferometer for High-Speed Flow Fluctuation Measurements

A multi-point focused laser differential interferometer (FLDI) has been developed to measure density fluctuations at 16 points along a line. A pair of cylindrical lenses on the transmitter side of a conventional single-point FLDI instrument form two closely spaced (≤200 μm), orthogonally polarized, parallel laser lines at the instrument’s focus. On the receiver side of the instrument, the interference of the beams on a 16-element photodiode array results in a single line of measurements. The further addition of a Nomarski prism creates two separate measurement lines, and the addition of a second photodiode array to the instrument enables simultaneous measurements of density fluctuations along the two lines separated by several millimeters. These two lines of measurement can be conveniently oriented at any azimuthal angle relative to the instrument’s optical axis on the measurement plane, coinciding with the instrument’s focus. Two experiments were performed to demonstrate the capabilities of the instrument. In the first experiment, a laser-induced breakdown spark generated a traveling spherical shock wave, and measurements of the resulting density disturbance and wave velocity were obtained. These results were compared to high-speed schlieren images of the shock wave acquired at 400 kHz. In the second experiment, the multi-point FLDI instrument was used to measure density disturbances in the boundary layer of a flat plate in a Mach 6 freestream flow. The measurements were made along two lines, both approximately 6 mm in length, extending from the surface of the plate through the boundary layer. High-speed schlieren images were acquired at 100 kHz during separate wind tunnel runs at matching unit Reynolds numbers to visualize the unsteady boundary layer flow and compare to the FLDI measurements.

Laser beam shaping↗

Analysis of the Amplitude Response of a Two-Point and a Two-Line/Multi-Point Focused Laser Differential Interferometer

Experiments were performed with two different focused laser differential interferometer (FLDI) instruments to assess the amplitude response of each instrument to a traveling shock-wave generated by a laser spark. The first FLDI instrument generated two measurement points, with each measurement point providing sensitivity to density fluctuations induced by the shockwave. By performing a cross-correlation between the signals obtained from each measurement point, the phase velocity of density fluctuations can be obtained. The second FLDI instrument generated two measurement lines that are oriented either parallel or orthogonal to the shockwave direction of travel with the resulting interference pattern sampled at multiple equally-spaced points along each line. As a result, this instrument provides density fluctuation measurement capability at multiple points simultaneously. When the measurement lines are oriented parallel to the shockwave direction of travel, the phase velocity, rate of change of the phase velocity, and acceleration of density fluctuations traveling along each line can be obtained by performing a cross-correlation between points along a single line. When the measurement lines are oriented orthogonal to the shockwave direction of travel, the spatially-varying phase velocity can be obtained by performing a cross-correlation between points at the same relative location on each measurement line. Computations of the shockwave generated by the laser spark are used to simulate the response of each instrument and are compared to the experimental results. High-speed Schlieren imaging has also been performed and is compared with the FLDI measurements and computational results.

Focused Laser Differential Interferometer↗

Two-Line Focused Laser Differential Interferometry of a Flat Plate Boundary Layer at Mach 6

Cylindrical optics have been added to a conventional two-point focused laser differential interferometry (FLDI) system to make measurements of density fluctuations at multiple points along two lines. Optics fixed in rotation mounts allow for the two lines to be oriented in any position relative to each other to make measurements of different physical phenomena. The system was first characterized in the laboratory using a laser-induced breakdown spark to provide a well-defined density fluctuation. The system was then installed at the NASA Langley 20 Inch Mach 6 Air Tunnel to make measurements through the boundary layer of a flat plate model and in the tunnel free stream.

Joshua M Weisberger↗

Two-Line Focused Laser Differential Interferometry of a Flat Plate Boundary Layer at Mach 6

Flat plate models are often used in wind tunnel testing to make measurements of the boundary layer, but preclude the use of conventional focused laser differential interferometry (FLDI) measurements, particularly for measurements at the surface of the model. By instead focusing light in only one direction using cylindrical optics, as opposed to two directions using only spherical optics, FLDI measurements can be made without truncating the beam at the model edges. Measurements at discrete points along a line can also be acquired, effectively acting as many individual FLDI instruments in a single measurement. Here, a two-line FLDI system has been constructed that allows for the measurement lines to be easily oriented at any azimuthal angle about the optical axis since the optics are housed in rotation mounts. Velocity measurements at points along a line or between two lines are also possible, since the beams are split similar to the two-point FLDI instrument. The system was first characterized in the laboratory using a laser-induced breakdown spark to provide a well-defined density fluctuation, along with high-speed schlieren measurements for comparison. The system was then installed at the NASA Langley 20-Inch Mach 6 Air Tunnel to make measurements through the boundary layer of a flat plate model and in the tunnel freestream.

Joshua M Weisberger↗

Linear Array Photodiode and Data Acquisition System Development for Multi-Point Line FLDI Measurements

The continuous development and advancement of multi-point line focused laser differential interferometry (FLDI) instruments requires larger numbers of acquisition channels to make more finely spatially resolved measurements of high-frequency flow fluctuations. No current off-the-shelf system exists for these line FLDI measurements, and so a single unified system consisting of a linear array photodiode, required electronics, and a data acquisition system has been developed to enhance the measurement resolution, both spatially and temporally. A printed circuit board was developed on which the linear array photodiode is mounted for inclusion into an existing FLDI system, and amplifier breakout boards incorporate electronics to condition the signal for the data acquisition system while retaining the necessary high frequency response. The data acquisition system uses a National Instruments chassis and multiple oscilloscope modules, with a specially-designed LabVIEW user interface for laboratory and wind tunnel setup, calibration, and testing. Details regarding the development of the linear array photodiode system and the data acquisition system are provided, and systematic testing of all components is presented, culminating in a test of the system integrated into an existing line FLDI system for measurements of a turbulent jet.

Joshua M Weisberger↗