A new, low power ionization source for elemental MS analysis of aqueous solutions is described. The liquid sampling-atmospheric pressure glow discharge (LSAPGD) operates by a process wherein the surface of the liquid emanating from a 75 μm i.d. glass capillary acts as the cathode of the direct current glow discharge. Analytecontaining solutions at a flow rate of 100 μL min-1 are vaporized by the passage of current, yielding gas phase solutes that are subsequently ionized in the < 5 W (maximum of 60 mA and 500 V), ~1 mm3 volume, plasma. The LS-APGD is mounted in place of the normal electrospray ionization source of a Thermo Scientific Exactive orbitrap mass spectrometer system. Basic operating characteristics are described, including the role of discharge power on mass spectral composition, the ability to obtain ultra-high resolution elemental isotopic patterns, and preliminary limits of detection attainable based on the injection of aliquots of multielement standards. While much optimization remains, it is believed that the LS-APGD may present a practical alternative to high-powered (>1 kW) plasma sources typically employed in elemental mass spectrometry, particularly for those cases where costs, operational overhead, and simplicity considerations are important.