Search NASA⌕ Search

Engineering topics

Behera, B.

Publications and source records attributed to Behera, B..

At least 19 records

Cryogenics and purification systems of the ICARUS T600 detector installation at Fermilab

This paper describes the cryogenic and purification systems of the ICARUS T600 detector in its present implementation at the Fermi National Laboratory, Illinois, U.S.A. The ICARUS T600 detector is made of four large Time Projection Chambers, installed in two separate containers of about 275 m 3 each. The detector uses liquid argon both as target and as active medium. For the correct operation of the detector, the liquid argon must be kept in very stable thermal conditions and the contamination of electronegative impurities must be consistently kept at the level of small fractions of parts per billion. The detector was previously operated in Italy, at the INFN Gran Sasso Underground Laboratory (LNGS), in a three-year run on the CERN to LNGS Long Baseline Neutrino Beam. For its operation on the Booster and NuMI neutrino beams at Fermilab, for the search of sterile neutrinos and measurements of neutrino-argon cross sections, the detector was moved from Gran Sasso to CERN for the upgrades required for operation at shallow depth with high intensity neutrino beams. The liquid argon containers, the thermal insulation and all the cryogenic equipment have been completely re-designed and rebuilt, following the schemes of the previous installation in Gran Sasso. The detector and all the equipment have been transported to Fermilab, where they have been installed, tested and recently put into operation. The work described in this paper has been conducted as a joint responsibility of CERN and Fermilab with the supervision provided by the ICARUS Collaboration. Design, installation, testing, commissioning and operation are the result of a common effort of CERN, Fermilab and INFN groups.

Cryogenic detectors↗

Spatial and temporal evaluations of the liquid argon purity in ProtoDUNE-SP

Liquid argon time projection chambers (LArTPCs) rely on highly pure argon to ensure that ionization electrons produced by charged particles reach readout arrays. ProtoDUNE Single-Phase (ProtoDUNE-SP) was an approximately 700-ton liquid argon detector intended to prototype the Deep Underground Neutrino Experiment (DUNE) Far Detector Horizontal Drift module. It contains two drift volumes bisected by the cathode plane assembly, which is biased to create an almost uniform electric field in both volumes. The DUNE Far Detector modules must have robust cryogenic systems capable of filtering argon and supplying the TPC with clean liquid. This paper will explore comparisons of the argon purity measured by the purity monitors with those measured using muons in the TPC from October 2018 to November 2018. A new method is introduced to measure the liquid argon purity in the TPC using muons crossing both drift volumes of ProtoDUNE-SP. For extended periods on the timescale of weeks, the drift electron lifetime was measured to be above 30 ms using both systems. A particular focus will be placed on the measured purity of argon as a function of position in the detector.

72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS↗

Design and implementation of the cosmic ray tagger system for the ICARUS detector at FNAL

The ICARUS-T600 Liquid Argon Time Projection Chamber is operating at Fermilab at shallow depth and thus exposed to a high flux of cosmic rays that can fake neutrino interactions. A cosmic ray tagging (CRT) system (∼ 1100 m 2 ), surrounding the cryostat with two layers of fiber embedded plastic scintillators, was developed to mitigate the cosmic ray induced background. Using nanosecond-level timing information, the CRT can distinguish incoming cosmic rays from outgoing particles from neutrino interactions in the TPC. In this paper an overview of the CRT system, its installation and commissioning at Fermilab, and its performance are discussed.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

The track-length extension fitting algorithm for energy measurement of interacting particles in liquid argon TPCs and its performance with ProtoDUNE-SP data

This paper introduces a novel track-length extension fitting algorithm for measuring the kinetic energies of inelastically interacting particles in liquid argon time projection chambers (LArTPCs). The algorithm finds the most probable offset in track length for a track-like object by comparing the measured ionization density as a function of position with a theoretical prediction of the energy loss as a function of the energy, including models of electron recombination and detector response. The algorithm can be used to measure the energies of particles that interact before they stop, such as charged pions that are absorbed by argon nuclei. The algorithm's energy measurement resolutions and fractional biases are presented as functions of particle kinetic energy and number of track hits using samples of stopping secondary charged pions in data collected by the ProtoDUNE-SP detector, and also in a detailed simulation. Additional studies describe the impact of the dE/dx model on energy measurement performance. The method described in this paper to characterize the energy measurement performance can be repeated in any LArTPC experiment using stopping secondary charged pions.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Calibration and simulation of ionization signal and electronics noise in the ICARUS liquid argon time projection chamber

The ICARUS liquid argon time projection chamber (LArTPC) neutrino detector has been taking physics data since 2022 as part of the Short-Baseline Neutrino (SBN) Program. This paper details the equalization of the response to charge in the ICARUS time projection chamber (TPC), as well as data-driven tuning of the simulation of ionization charge signals and electronics noise. The equalization procedure removes non-uniformities in the ICARUS TPC response to charge in space and time. This work leverages the copious number of cosmic ray muons available to ICARUS at the surface. The ionization signal shape simulation applies a novel procedure that tunes the simulation to match what is measured in data. The end result of the equalization procedure and simulation tuning allows for a comparison of charge measurements in ICARUS between Monte Carlo simulation and data, showing good performance with minimal residual bias between the two.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Angular dependent measurement of electron-ion recombination in liquid argon for ionization calorimetry in the ICARUS liquid argon time projection chamber

This paper reports on a measurement of electron-ion recombination in liquid argon in the ICARUS liquid argon time projection chamber (LArTPC). A clear dependence of recombination on the angle of the ionizing particle track relative to the drift electric field is observed. An ellipsoid modified box (EMB) model of recombination describes the data across all measured angles. These measurements are used for the calorimetric energy scale calibration of the ICARUS TPC, which is also presented. The impact of the EMB model is studied on calorimetric particle identification, as well as muon and proton energy measurements. Accounting for the angular dependence in EMB recombination improves the accuracy and precision of these measurements.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

DUNE Phase II: scientific opportunities, detector concepts, technological solutions

The international collaboration designing and constructing the Deep Underground Neutrino Experiment (DUNE) at the Long-Baseline Neutrino Facility (LBNF) has developed a two-phase strategy toward the implementation of this leading-edge, large-scale science project. The 2023 report of the US Particle Physics Project Prioritization Panel (P5) reaffirmed this vision and strongly endorsed DUNE Phase I and Phase II, as did the European Strategy for Particle Physics. While the construction of the DUNE Phase I is well underway, this White Paper focuses on DUNE Phase II planning. DUNE Phase-II consists of a third and fourth far detector (FD) module, an upgraded near detector complex, and an enhanced 2.1 MW beam. The fourth FD module is conceived as a “Module of Opportunity”, aimed at expanding the physics opportunities, in addition to supporting the core DUNE science program, with more advanced technologies. This document highlights the increased science opportunities offered by the DUNE Phase II near and far detectors, including long-baseline neutrino oscillation physics, neutrino astrophysics, and physics beyond the standard model. It describes the DUNE Phase II near and far detector technologies and detector design concepts that are currently under consideration. A summary of key R&D goals and prototyping phases needed to realize the Phase II detector technical designs is also provided. DUNE's Phase II detectors, along with the increased beam power, will complete the full scope of DUNE, enabling a multi-decadal program of groundbreaking science with neutrinos.

72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS↗

First measurement of the total inelastic cross section of positively charged kaons on argon at energies between 5.0 and 7.5 GeV

ProtoDUNE Single-Phase (ProtoDUNE-SP) is a 770-ton liquid argon time projection chamber that operated in a hadron test beam at the CERN Neutrino Platform in 2018. We present a measurement of the total inelastic cross section of charged kaons on argon as a function of kaon energy using 6 and 7 GeV/𝑐 beam momentum settings. The flux-weighted average of the extracted inelastic cross section at each beam momentum setting was measured to be 380 ± 26 mbarns for the 6 GeV/𝑐 setting and 379 ± 35 mbarns for the 7 GeV/𝑐 setting.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Characterization and novel application of power over fiber for electronics in a harsh environment

Power-over-Fiber (PoF) technology has been used extensively in settings where high voltages require isolation from ground. In a novel application of PoF, power is provided to photon detector modules located on a surface at ∼ 300 kV with respect to ground in the planned DUNE experiment. In cryogenic environments, PoF offers a reliable means of power transmission, leveraging optical fibers to transfer optical power. PoF technology excels in maintaining low noise levels when delivering power to sensitive electronic systems operating in extreme temperatures and high voltage environments. This paper presents the R&D effort of PoF in extreme conditions and underscores its capacity to revolutionize power delivery and management in critical applications, offering a dependable solution with low noise, optimal efficiency (∼ 51%), and superior isolation.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Scintillation light in SBND: simulation, reconstruction, and expected performance of the photon detection system

SBND is the near detector of the Short-Baseline Neutrino program at Fermilab. Its location near to the Booster Neutrino Beam source and relatively large mass will allow the study of neutrino interactions on argon with unprecedented statistics. This paper describes the expected performance of the SBND photon detection system, using a simulated sample of beam neutrinos and cosmogenic particles. Its design is a dual readout concept combining a system of 120 photomultiplier tubes, used for triggering, with a system of 192 X-ARAPUCA devices, located behind the anode wire planes. Furthermore, covering the cathode plane with highly-reflective panels coated with a wavelength-shifting compound recovers part of the light emitted towards the cathode, where no optical detectors exist. We show how this new design provides a high light yield and a more uniform detection efficiency, an excellent timing resolution and an independent 3D-position reconstruction using only the scintillation light. Finally, the whole reconstruction chain is applied to recover the temporal structure of the beam spill, which is resolved with a resolution on the order of nanoseconds.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Power over fiber development for HEP detectors

Power-over-Fiber (PoF) technology has been used extensively in settings where high voltages require isolation from ground and electromagnetic isolation is critical. In cryogenic environments, PoF offers a reliable power transmission technology, leveraging optical fibers to transfer power with minimal system degradation. PoF technology excels in maintaining low noise levels and isolation when delivering power to sensitive electronic systems operating in extreme temperature ranges and high voltage environments. Here, in a novel application of PoF for a HEP detector, power is provided to photon detector modules located on a surface at ~300 kV with respect to ground in the planned DUNE experiment. This summary paper of the PoF talk at the 16th PISA Meeting on Advanced Detectors highlights the R&D effort of PoF in extreme conditions and underscores its capacity to revolutionize power delivery and management in critical applications offering a dependable solution with low noise, optimal efficiency, and superior isolation. The DUNE (Abi et al., 2020) experiment will soon deploy large liquid argon (LAr) time projection chambers (TPC) to detect neutrino interactions and other particle physics phenomena. In addition to the particle tracking provided by the TPC, photon detectors, powered by a first ever PoF system, in the cryostat will leverage the high scintillation light yield of LAr to provide crucial timing and additional calorimetric information.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Performance of a Modular Ton-Scale Pixel-Readout Liquid Argon Time Projection Chamber

The Module-0 Demonstrator is a single-phase 600 kg liquid argon time projection chamber operated as a prototype for the DUNE liquid argon near detector. Based on the ArgonCube design concept, Module-0 features a novel 80k-channel pixelated charge readout and advanced high-coverage photon detection system. In this paper, we present an analysis of an eight-day data set consisting of 25 million cosmic ray events collected in the spring of 2021. We use this sample to demonstrate the imaging performance of the charge and light readout systems as well as the signal correlations between the two. We also report argon purity and detector uniformity measurements and provide comparisons to detector simulations.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

The DUNE Far Detector Vertical Drift Technology. Technical Design Report

DUNE is an international experiment dedicated to addressing some of the questions at the forefront of particle physics and astrophysics, including the mystifying preponderance of matter over antimatter in the early universe. The dual-site experiment will employ an intense neutrino beam focused on a near and a far detector as it aims to determine the neutrino mass hierarchy and to make high-precision measurements of the PMNS matrix parameters, including the CP-violating phase. It will also stand ready to observe supernova neutrino bursts, and seeks to observe nucleon decay as a signature of a grand unified theory underlying the standard model. The DUNE far detector implements liquid argon time-projection chamber (LArTPC) technology, and combines the many tens-of-kiloton fiducial mass necessary for rare event searches with the sub-centimeter spatial resolution required to image those events with high precision. The addition of a photon detection system enhances physics capabilities for all DUNE physics drivers and opens prospects for further physics explorations. Given its size, the far detector will be implemented as a set of modules, with LArTPC designs that differ from one another as newer technologies arise. In the vertical drift LArTPC design, a horizontal cathode bisects the detector, creating two stacked drift volumes in which ionization charges drift towards anodes at either the top or bottom. The anodes are composed of perforated PCB layers with conductive strips, enabling reconstruction in 3D. Light-trap-style photon detection modules are placed both on the cryostat's side walls and on the central cathode where they are optically powered. This Technical Design Report describes in detail the technical implementations of each subsystem of this LArTPC that, together with the other far detector modules and the near detector, will enable DUNE to achieve its physics goals.

47 OTHER INSTRUMENTATION↗

Doping liquid argon with xenon in ProtoDUNE Single-Phase: effects on scintillation light

Doping of liquid argon TPCs (LArTPCs) with a small concentration of xenon is a technique for light-shifting and facilitates the detection of the liquid argon scintillation light. In this paper, we present the results of the first doping test ever performed in a kiloton-scale LArTPC. From February to May 2020, we carried out this special run in the single-phase DUNE Far Detector prototype (ProtoDUNE-SP) at CERN, featuring 720 t of total liquid argon mass with 410 t of fiducial mass. A 5.4 ppm nitrogen contamination was present during the xenon doping campaign. The goal of the run was to measure the light and charge response of the detector to the addition of xenon, up to a concentration of 18.8 ppm. The main purpose was to test the possibility for reduction of non-uniformities in light collection, caused by deployment of photon detectors only within the anode planes. Light collection was analysed as a function of the xenon concentration, by using the pre-existing photon detection system (PDS) of ProtoDUNE-SP and an additional smaller set-up installed specifically for this run. In this paper we first summarize our current understanding of the argon-xenon energy transfer process and the impact of the presence of nitrogen in argon with and without xenon dopant. We then describe the key elements of ProtoDUNE-SP and the injection method deployed. Two dedicated photon detectors were able to collect the light produced by xenon and the total light. The ratio of these components was measured to be about 0.65 as 18.8 ppm of xenon were injected. We performed studies of the collection efficiency as a function of the distance between tracks and light detectors, demonstrating enhanced uniformity of response for the anode-mounted PDS. We also show that xenon doping can substantially recover light losses due to contamination of the liquid argon by nitrogen.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Time calibration and synchronization of the scintillation light detection system in ICARUS-T600

The ICARUS-T600 Liquid Argon (LAr) Time Projection Chamber (TPC) is presently taking data in the Short Baseline Neutrino (SBN) program at Fermilab (U.S.A.) to search for a possible LSND-like sterile neutrino signal at Δm 2 ≈ 1 eV 2 with the Booster Neutrino Beam (BNB). A light detection system, based on 360 large area Photo-Multiplier Tubes (PMTs), has been realized for ICARUS-T600 to detect VUV photons produced by the passage of ionizing particles in LAr. This system is fundamental for the TPC operation, providing an efficient trigger and contributing to the 3D reconstruction of events. Moreover, since the detector is exposed to a huge flux of cosmic rays due to its shallow depths installation, the light detection system allows for the time reconstruction of events, contributing to the identification and to the selection of genuine neutrino interactions. The correct time reconstruction of events requires the precise knowledge of the delay of each PMT channel and a good synchronization of recording electronics, this last based on fast sampling digitizers. To achieve a time resolution better than 1 ns, we perform three consecutive timing corrections deployed at different stages of the optical data flow. Results demonstrate the capability of the ICARUS-T600 light detection system to allow a precise reconstruction of the temporal evolution of each event occurring in the detector and the association of neutrino events with the bunched structure of BNB.

72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS↗

Time calibration and synchronization of the scintillation light detection system in ICARUS-T600

The ICARUS-T600 Liquid Argon (LAr) Time Projection Chamber (TPC) is presently used as a far detector of the Short Baseline Neutrino (SBN) program at Fermilab (USA) to search for a possible LSND-like sterile neutrino signal at $\Delta m^2 \approx 1 eV^2$ with the Booster Neutrino Beam (BNB). A light detection system, based on 360 large area Photo-Multiplier Tubes (PMTs), has been realized for ICARUS-T600 to detect VUV photons produced after the passage of ionizing particles in LAr. This system is fundamental for the TPC operation, providing an efficient trigger and contributing to the 3D reconstruction of events. Moreover, since the detector is exposed to a huge flux of cosmic rays due to its shallow depths operations, the light detection system allows for the time reconstruction of events, contributing to the identification and to the selection of neutrino interactions within the BNB spill gate. The correct time reconstruction of events requires the precise knowledge of the delay of each PMT channel and a good synchronization of recording electronics, this last based on fast sampling digitizers.

72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS↗

ICARUS at the Fermilab Short-Baseline Neutrino program: initial operation

The ICARUS collaboration employed the 760-ton T600 detector in a successful 3-year physics run at the underground LNGS laboratory, performing a sensitive search for LSND-like anomalous $\nu _e$ appearance in the CERN Neutrino to Gran Sasso beam, which contributed to the constraints on the allowed neutrino oscillation parameters to a narrow region around 1 eV$^2$. After a significant overhaul at CERN, the T600 detector has been installed at Fermilab. In 2020 the cryogenic commissioning began with detector cool down, liquid argon filling and recirculation. ICARUS then started its operations collecting the first neutrino events from the booster neutrino beam (BNB) and the Neutrinos at the Main Injector (NuMI) beam off-axis, which were used to test the ICARUS event selection, reconstruction and analysis algorithms. ICARUS successfully completed its commissioning phase in June 2022. The first goal of the ICARUS data taking will be a study to either confirm or refute the claim by Neutrino-4 short-baseline reactor experiment. ICARUS will also perform measurement of neutrino cross sections with the NuMI beam and several Beyond Standard Model searches. After the first year of operations, ICARUS will search for evidence of sterile neutrinos jointly with the Short-Baseline Near Detector, within the Short-Baseline Neutrino program. In this paper, the main activities carried out during the overhauling and installation phases are highlighted. Preliminary technical results from the ICARUS commissioning data with the BNB and NuMI beams are presented both in terms of performance of all ICARUS subsystems and of capability to select and reconstruct neutrino events.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗