A technique has been developed to automatically correct for drifts in the radiometric sensitivity of the detector channels in a direct-reading emission spectrometer. The method utilizes a 1000 W tungsten-halogen reference lamp to illuminate the detectors through the same optical path as that traversed during the analysis of the sample. Detector channel responses to the light are compared to those for the same light intensity at the time of analytical calibration. This corrects for the drift. It is noted that with the exception of positioning the lamp, the procedure is fully automatic.

Chapman, G. B., II↗

Development of a drift-correction procedure for a direct-reading spectrometer

A procedure which provides automatic correction for drifts in the radiometric sensitivity of each detector channel in a direct-reading emission spectrometer is described. Such drifts are customarily controlled by the regular analyses of standards, which provide corrections for changes in the excitational, optical, and electronic components of the instrument. This standardization procedure, however, corrects for the optical and electronic drifts. It is a step that must be taken if the time, effort, and cost of processing standards is to be minimized. This method of radiometric drift correction uses a 1,000-W tungsten-halogen reference lamp to illuminate each detector through the same optical path as that traversed during sample analysis. The responses of the detector channels to this reference light are regularly compared with channel response to the same light intensity at the time of analytical calibration in order to determine and correct for drift. Except for placing the lamp in position, the procedure is fully automated and compensates for changes in spectral intensity due to variations in lamp current. A discussion of the implementation of this drift-correction system is included.

Chapman, G. B., II↗

Sample preparation of metal alloys by electric discharge machining

Electric discharge machining was investigated as a noncontaminating method of comminuting alloys for subsequent chemical analysis. Particulate dispersions in water were produced from bulk alloys at a rate of about 5 mg/min by using a commercially available machining instrument. The utility of this approach was demonstrated by results obtained when acidified dispersions were substituted for true acid solutions in an established spectrochemical method. The analysis results were not significantly different for the two sample forms. Particle size measurements and preliminary results from other spectrochemical methods which require direct aspiration of liquid into flame or plasma sources are reported.

Chapman, G. B., II↗

Determination of work functions near melting points of refractory metals by using a direct-current arc.

A direct-current arc in argon at atmospheric pressure was used to determine effective work functions of refractory metals, including tantalum, tungsten, molybdenum, and niobium. The procedure is experimentally advantageous, because surface cleanliness of the specimen is not critical, high vacuum is not required, and the anode-cathode spacing is not critical. The experimental procedure involves striking an arc to a metal wire cathode to form a melted ball having an emitting area defined by its diameter. The literature melting point of the metal is taken as the emitting temperature. By using these parameters and the known arc current, effective work functions were calculated from the Richardson-Dushman equation. The calculated work functions agree with recommended handbook values to within about 0.1 V and have typical repeatabilities of 0.02 V.

Gordon, W. A.↗

Determination of work functions near melting points of refractory metals by using a direct-current arc

Effective work functions of refractory metals at temperatures near their melting points were determined by using a direct-current arc. A metal wire connected as the cathode was melted by striking an arc discharge in an argon atmosphere. A melted sphere was formed with a definite emitting area which was calculated from the sphere diameter measured after terminating the arc. Effective work functions were calculated from the Richardson-Dushman equation by using this emission area. The procedure is experimentally advantageous because surface cleanliness of the specimen is not critical, high vacuum is not required, and the anode-cathode spacing is not critical.

Gordon, W. A.↗