BioID/LID System Cartridge Testing Report
LLNL produced two lots of lyophilized LAMP assay containing cartridges, LNL022818 and LNL071818, for testing on Gen3 BioID systems. Thirty-nine cartridges were tested for lot LNL022818. All thirty-nine cartridges performed as expected regarding correct positive control (IPC) and negative control (NTC) reactions. Five different templates were tested on these 39 cartridges. The BioID system detected 79.1% true positives and 9.3% false positives along with 98.9% true negatives and 2.9% false negatives for cartridge lot LNL022818. Thirty cartridges were tested for lot LNL071818. It was noted early that due to a manufacturing issue with the cartridges related to incomplete sonic welding of the cartridge, a large fraction of lot LNL071818 cartridges leaked. Insufficient funds were available to replace this lot of cartridges. Twenty three of 30 cartridges performed as expected regarding correct positive control (IPC) and negative control (NTC) reactions. Four different templates were tested on these 30 cartridges. The BioID system detected 77% true positives and 23% false positives along with 96.9% true negatives and 3.1% false negatives for cartridge lot LNL071818. For all tested cartridges using DNA template, DNA was added at a final concentration of 25 pg/ul. All reactions contained positive control DNA at 2.5 pg/ul.