New Design Yields Robust Large-Area Framing Camera

Framing cameras provide a discrete series of images over a short period of time in a manner that mimics a high-speed movie camera, but with individual shutter times that must be extremely short in order to capture freeze-frame images of rapidly evolving phenomena such as high-explosive shock-driven ejecta, dynamic compression of metals, and high-velocity fluid flow. The NNSS has designed and fielded a new, large-area, gated framing camera called Kraken. The camera design emphasized manufacturability and flexibility by improving imager yield and creating a camera architecture to shorten design cycles. Design strategies and field data are presented.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Kraken Camera Development and Fielding Overview [Slides]

Presentation given at a LANL radiography workshop held April 6, 2021, at LANL LANSCE. This presentation is a shortened version of an earlier presentation made on the Kraken camera (DOE/NV/03624--0924) with some additional content.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

New Design Yields Robust Large-Area Framing Camera

Manuscript to be submitted to Review of Scientific Instruments to determine interest in possible publication. Website: https://aip.scitation.org/rsi/authors/manuscript High-speed imaging is a core diagnostic capability for the Nevada National Security Site (NNSS). Data from experiments that measure the behavior of materials under extreme conditions are essential for the stockpile mission. For example, during subcritical experiments, materials that are shocked and compressed under high-explosive drive change behavior rapidly over timescales measured in millionths or even billionths of a second. Characterizing and understanding this behavior requires radiographic and/or optical imaging to track motion on these timescales. Framing cameras provide a discrete series of images over a short period of time in a manner that mimics a high-speed movie camera, but with individual shutter times that must be extremely short in order to capture freeze-frame images of rapidly evolving phenomena such as high-explosive shock-driven ejecta, dynamic compression of metals, and high-velocity fluid flow.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Kraken Imager and Camera Systems

Presentation describes the Kraken imager and camera systems under development for use at the Nevada National Security Site for work conducted by the National Nuclear Security Administration.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

FY19 L2 MRT 6744 Milestone Completion: Digital Time-Resolved Spot Diagnostic (D-TRSD)

This milestone report describes development of the Digital Time-Resolved Spot Diagnostic (D-TRSD) into a functioning diagnostic that has been proven in the laboratory. Laboratory tests complete the MRT 6744 exit criteria with the recording and analysis of a spot size measurement for two consecutive pulses on a two-pulse source. Additional channels will be added in the near future to provide an 84-channel measurement capability. The new modular design is more portable than the previous generations and will be much easier to set up at radiographic facilities across the complex.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Digital Time-Resolved Spot Diagnostic

The digital time-resolved spot diagnostic (D-TRSD) is intended to be a fast-response, time-dependent beam diagnostic for measuring the spot size emitted from radiographic x-ray sources. It will measure spot sizes from submillimeters up to several millimeters for photon energies in the range of 150 keV to 20 MeV with doses as low as 0.03 rad up to 500 rad. The D-TRSD performs fast data acquisition as radiographic x-ray sources typically produce pulses in the tens of nanosecond ranges. The system will be optimized to capture multiple individual pulses in time within the same data record. The data collected will be used to support beam tuning and source optimization with on-the-fly evaluation of radiographic source performance. The diagnostic is based on penumbral imaging techniques. A scintillating fiber array of parallel aligned fibers is positioned in the beam path with a high-Z rolled edge inserted between the source and the array, thereby creating the penumbral x-ray signal that is captured by the D-TRSD system. The light from the scintillation conversion is recorded by individual avalanche photodiode detectors integrated into custom digitizer electronics. The penumbra is used to calculate the spot of the beam on the axis orthogonal to the fiber array based on the magnification from the rolled edge, the resolution of the spaced fibers, and the rotation of the array normal to the beam path. Many arrays can be used to capture additional spot widths about different rotations. The D-TRSD system is modular, software-driven, and easier to use than legacy TRSD systems.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Digital Time-Resolved Spot Diagnostic

The digital time-resolved spot diagnostic (D-TRSD) is a fast-response, time-dependent beam diagnostic for measuring the emitted spot size and beam target interactions from radiographic x-ray sources. Legacy streaked TRSD analysis requires image processing and data extraction, which give noisy results. Processed data are used to extract edge spread and corresponding line spread functions. Digital TRSD simplifies this process, eliminating the complexity of streak camera–based systems.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

All-Digital Time-Resolved Spot Diagnostic

Presentation describes research into construction of a time-resolved spot diagnostic using mostly commercial off-the-shelf components, with system requirements to include (among others) comparable or better signal-to-noise ratio than a streak camera; low-loss fiber network, likely less than a few dB; optimized interface to the scintillators; and fast detectors to capture the pulse shape, 500 MHz to 1 GHz analog bandwidth with good linearity.