Results of dynamic testing of GP-B spherical gyroscopes

Laboratory tests of the spherical electrostatically levitated cryogenically cooled coated gyroscope being developed for the Gravity Probe B (GP-B) spacecraft (Bardas et al., 1986) are reported. Spin speed and the dc components of the trapped magnetic field are measured with three orthogonal pickup loops attached to SQUID detectors as the levitated gyro is brought up to speed by an He gas jet. Data on the spin-vector time history, mass unbalance, higher rotor-shape harmonics, and spin-vector position are presented in extensive graphs and characterized in detail, and a mathematical model of the electrostatic suspension torques is derived. Prototype gyro 86-4 is found to have mass unbalance within the range required for the GP-B mission (to detect the geodetic and motional effects predicted by general relativity theory).

Keiser, G. M.↗

Mass unbalance analysis for GP-B rotors

Using three different approaches, laboratory results were obtained for the mass unbalance of the NASA Gravity Probe B rotors designed to test Einstein's gravitational theory. In the first method, measurements of the precession rate of the spinning rotor are deduced from spin-frequency components of the trapped flux. The other approaches involve the measuring of niobium coating thickness using an electron back-scattering device and the measuring of the periodicities of a tumbling rotor. Assumptions of the methods are discussed, along with reasons for discrepancies among the results.