Spectroscopic diagnostics of high-temperature plasma in stellar coronae using Fe xxiv – xxiv K-shell lines with XRISM
Explore the source record for details and available documents.
Engineering topics
Publications and source records attributed to Frederick S Porter.
Explore the source record for details and available documents.
The Hitomi results for the Perseus cluster have shown that accurate atomic models are essential to the success of X-ray spectroscopic missions and just as important as the store of knowledge on instrumental calibration and astrophysical modeling. Preparing the models requires a multifaceted approach, including theoretical calculations, laboratory measurements, and calibration using real observations. In a previous paper, we presented a calculation of the electron impact cross sections on the transitions forming the Fe-L complex. In the present work, we systematically tested the calculation against cross-sections of ions measured in an electron beam ion trap experiment. A two-dimensional analysis in the electron beam energies and X-ray photon energies was utilized to disentangle radiative channels following dielectronic recombination, direct electron-impact excitation, and resonant excitation processes in the experimental data. The data calibrated through laboratory measurements were further fed into a global modeling of the Chandra grating spectrum of Capella. We investigated and compared the fit quality, as well as the sensitivity of the derived physical parameters to the underlying atomic data and the astrophysical plasma modeling. We further list the potential areas of disagreement between the observations and the present calculations, which, in turn, calls for renewed efforts with regard to theoretical calculations and targeted laboratory measurements.
The X-ray Integral Field Unit (X-IFU) instrument on the Athena mission will be positioned at the Lagrangian point L2 and be subject to cosmic rays generated by astrophysics sources, primarily relativistic protons. Previous simulations have shown that particles of energy higher than 150 MeV will make it through the outer layers of the satellite. They will reach the detector wafer with a rate of 3 cts/sq. cm/s and a most probable energy deposited in the Si frame supporting the array at 150 keV. These events can affect the energy resolution of the detectors through the thermal fluctuations that they produce. This study assesses this potential problem and discusses two suggested design approaches to decrease the impact of cosmic ray in order to limit their effect to their allocation of 0.2 eV within the Athena/X-IFU energy-resolution budget. The first is the addition of a coating layer of high heat capacity material (e.g., Pd) and the second is the splitting of this coating into two thermal regions near the TES array to keep the heat away from the array. Implementing these two features is predicted to cause a decrease in the number of events above 1 μK by more than a factor 10 to ~ 1.5 cps when compared to an equivalent design without these features.
Arrays of high-density microcalorimeters require careful heat sinking in order to minimize the thermal crosstalk between nearby pixels. For the array of microcalorimeters developed for the Athena X-ray Integral Field Unit instrument, which has more than 3000 pixels on a 275 µm pitch, it is essential to address this problem in order to meet the energy-resolution requirements. The instrument’s energy-resolution budget requires that the impact of the thermal crosstalk on the energy resolution be a contribution that, added in quadrature to other energy-resolution contributions, is less than 0.2 eV. This value results in a derived requirement that the ratio between the amplitude of the crosstalk signal to an X-ray pulse (for example at 6 keV) is less than 1 × 10(exp −3) (for the first neighbor), less than 4 × 10(exp −4) (for the diagonal neighbor) and less than 8 × 10(exp −5) (for the second nearest neighbor). We have measured the thermal crosstalk levels between pixels in various geometries and configurations. The results show a crosstalk ratio which is at least a factor of 4 lower than the derived requirement. We also developed a finite element (FEM) 2D thermal model to predict the thermal behavior of large-scale arrays. This model successfully simulates the measured data in terms of pulse amplitude and time constants.
X‐ray astronomy and ground‐based atomic physics have a long history of fruitful collaboration: Sound understanding of the underlying atomic physics is the key to reliable interpretation of the spectra from celestial sources; conversely, astronomical spectra have been used to benchmark and advance atomic physics. This interplay is about to become even more important as we enter a new era of high‐resolution X‐ray astrophysics with large effective collection area. Although high‐resolution observations with the gratings on the Chandra and XMM‐Newton observatories continue to drive new science, upcoming planned and proposed missions will open up new discovery space in the near future that is currently challenging to access: high‐resolution spectroscopy on extended sources, in the Fe K band, and on short time scales. This review summarizes open questions in these areas and the design parameters for the Hitomi, XRISM, Athena, and Arcus observatories. The expected high quality of spectra taken with these observatories puts new constraints on the accuracy of atomic reference data required to take full advantage of the diagnostic potential of these spectra.
The Soft X-ray Spectrometer (SXS) instrument[1] on Astro-H[2] will use a 3-stage ADR[3] to cool the microcalorimeter array to 50 mK. In the primary operating mode, two stages of the ADR cool the detectors using superfluid helium at ≤1.20 K as the heat sink[4]. In the secondary mode, which is activated when the liquid helium is depleted, the ADR uses a 4.5 K Joule-Thomson cooler as its heat sink. In this mode, all three stages operate together to continuously cool the (empty) helium tank and singleshot cool the detectors. The flight instrument - dewar, ADR, detectors and electronics - were integrated in 2014 and have since undergone extensive performance testing. This paper presents a thermodynamic analysis of the ADR's operation, including cooling capacity, heat rejection to the heat sinks, and various measures of efficiency.
A 3-stage adiabatic demagnetization refrigerator (ADR) is used on the Soft X-ray Spectrometer instrument on Astro-H to cool a 6x6 array of x-ray microcalorimeters to 50 mK. The ADR is supported by a cryogenic system consisting of a superfluid helium tank, a 4.5 K Joule-Thomson (JT) cryocooler, and additional 2-stage Stirling cryocoolers that pre-cool the JT cooler and cool radiation shields within the cryostat. The ADR is configured so that it can use either the liquid helium or the JT cryocooler as its heat sink, giving the instrument an unusual degree of tolerance for component failures or degradation in the cryogenic system. The flight detector assembly, ADR and dewar were integrated into the flight dewar in early 2014, and have since been extensively characterized and calibrated. This paper summarizes the operation and performance of the ADR in all of its operating modes
The measured noise in transition-edge sensor (TES) microcalorimeters with a Mo/Au bilayer has been shown to have a significant contribution from internal thermal fluctuation noise (ITFN) between the absorber and the TES. Previous measurements showed the relevant thermal link leading to the noise was the thermal conductance of the TES bilayer, and ITFN has been shown to correlate with the sheet resistance of the TES. As a result, ITFN is particularly significant in devices with a large bilayer sheet resistance. However, because of more favorable transition shapes, such high resistance devices have recently been shown to be preferable under alternating current bias, which is required for the frequency division multiplexed readout baselined for the X-ray Integral Field Unit (X-IFU) instrument on the ATHENA satellite. Therefore, it is important to investigate how else to minimize the ITFN beyond changes to the sheet resistance. In this presentation, we will discuss our investigation of how changes to the TES design can affect ITFN for a fixed sheet resistance. We will present how ITFN is altered by changes to the aspect ratio of the TES, the addition of normal metal banks parallel to the TES current direction, and the size and position of the connections between the absorber and the TES. We will also discuss how these design changes impact the superconducting transition shape, and the thermal conductance from the TES to the thermal bath. These results may provide routes for further improvements of TES performance, and a deeper understanding of the physics of the thermal network in a TES microcalorimeter.
The Resolve instrument on the X-Ray Imaging and Spectroscopy Mission (XRISM) uses an Adiabatic Demagnetization Refrigerator (ADR) to cool extremely sensitive x-ray microcalorimeters to 50 mK. A 40-liter superfluid helium dewar and a 4.5 K Joule Thomson (JT)/Stirling cryocooler serve as heat sinks for the ADR and staged cooling for the dewar. The superfluid helium lifetime is required to last at least 3 years on orbit with margin. The superfluid helium dewar developed a small leak (~5e-7 Pa m^3/s) at low temperature which compromised the dewar vacuum in two days of ground testing. It turns out that this leak disappears completely at room temperature (<5e-13 Pa m^3/s). This presentation describes the lengthy process of characterizing, isolating, and ultimately, repairing this leak.
The Advanced Telescope for High ENergy Astrophysics (ATHENA) mission requires high quantum efficiency (QE) x-ray absorption, >90.6% at 7 keV and low specific heat capacity, 0.731 pJ/K. The designed ATHENA x-ray absorbers are cantilevered square tiles (pitch of 317 microns) of 1.05 μm thick Au and 5.51 μm thick Bi electroplated films supported by stems that connect the absorber to the detector below. We discuss some of the methods used to produce x-ray absorbers meeting these specifications for ATHENA. To tune the thermal conductance of the device and the effect on the normal-to superconducting transition shape, the stems need to be small diameter and can have a weak bottle-neck connection to the substrate. A funnel shape of the stem using a proximity exposed photoresist mold has been developed to improve the strength of the connections. Further requirements on the absorbers include low levels of fine particulate remaining on the substrate after production and zero shorts between absorbers due to incomplete ion milling or trapped fine particular between absorbers. To optimize for post patterning substrate cleanliness and absorber yield, we have examined several methods of absorber fabrication. Three such methods are 1) an ion mill/wet etch combination, 2) a photoresist mold for electroplating followed by wet or dry etch to remove the seed layer, and 3) an electroplating process with leveling to smooth the surface followed by ion mill to separate the absorbers. The different combinations of wet and dry etches lead to different yields and surface appearance of the absorber sidewalls. We present results on the achieved pixel yields and the energy resolution of pixels made with the various fabrication methods. We discuss the impact of absorber patterning method on performance, uniformity, and yield.
The Athena X-Ray Integral Field Unit (X-IFU) comprises a 2376-pixel array of transition edge sensors (TES) read out with time-division multiplexing (TDM). X-IFU will provide spatially resolved, high-resolution spectroscopy (2.5 eV full-width-half-maximum up to 7 keV) over the energy range 0.2 to 12 keV, with an absolute energy scale accuracy of 0.4 eV. The energy scale function maps the optimally filtered pulse height, in arbitrary engineering units, to real calibrated energy. Uncertainties in the calibration can result from imperfect fitting of the energy scale between the known calibration points. Furthermore, temporal changes in the TES operating environment, such as heat-sink temperature, magnetic field and bias voltage, can cause significant variations in the detector gain function over time. If not properly corrected, this can result in degradation of the energy resolution, and systematic errors in the absolute energy scale. The non-linear nature of TES detectors, coupled with the possibility of multiple simultaneously occurring sources of drift, can make effective corrections over the full bandpass of the instrument extremely challenging. Athena X-IFU will employ an on-board calibration source that provides known reference x-ray lines. This provides real-time monitoring of the gain stability of the detector system and information that can be used to correct for gain drifts. For X-IFU the baseline approach is to measure a series of calibration curves under different environmental conditions, which bound the expected drifts the instrument is predicted to see over the course of the mission. Using the information from the in-flight calibration source, these energy scale functions can be interpolated to generate a new corrected energy scale as a function of time. In this paper we discuss progress towards demonstrating that the X-IFU energy scale requirements can be met. We present measurements on ~ 200 pixels in a prototype X-IFU array read out with 8-column x 32-row TDM. We use a rotating target source containing 12 fluorescent targets to generate x-ray lines covering the energy range 4 keV (Sc-Kα) to 12 keV (Br-Kα). We present measurements of the non-linear energy scale function and show how variations in heat-sink temperature, TES bias voltage and magnetic field affect the shape of TES energy scale differently and introduce different residual gain errors over the bandpass. We explore different drift correction algorithms that use either a single or multiple referential lines to track and correct the gain from these various sources of drift. In addition to the pulse-height, the DC ‘baseline’ level of the TES can contain information about its bias conditions. Thus, we test a multi-parameter gain correction algorithm that attempts to incorporate both the pulse height and the additional baseline information into the algorithm.