SU-F-T-201: Acceleration of Dose Optimization Process Using Dual-Loop Optimization Technique for Spot Scanning Proton Therapy

Purpose: The purpose was to demonstrate a developed acceleration technique of dose optimization and to investigate its applicability to the optimization process in a treatment planning system (TPS) for proton therapy. Methods: In the developed technique, the dose matrix is divided into two parts, main and halo, based on beam sizes. The boundary of the two parts is varied depending on the beam energy and water equivalent depth by utilizing the beam size as a singular threshold parameter. The optimization is executed with two levels of iterations. In the inner loop, doses from the main part are updated, whereas doses from the halo part remain constant. In the outer loop, the doses from the halo part are recalculated. We implemented this technique to the optimization process in the TPS and investigated the dependence on the target volume of the speedup effect and applicability to the worst-case optimization (WCO) in benchmarks. Results: We created irradiation plans for various cubic targets and measured the optimization time varying the target volume. The speedup effect was improved as the target volume increased, and the calculation speed increased by a factor of six for a 1000 cm3 target. An IMPT plan for the RTOG benchmark phantom was created in consideration of ±3.5% range uncertainties using the WCO. Beams were irradiated at 0, 45, and 315 degrees. The target’s prescribed dose and OAR’s Dmax were set to 3 Gy and 1.5 Gy, respectively. Using the developed technique, the calculation speed increased by a factor of 1.5. Meanwhile, no significant difference in the calculated DVHs was found before and after incorporating the technique into the WCO. Conclusion: The developed technique could be adapted to the TPS’s optimization. The technique was effective particularly for large target cases.

60 APPLIED LIFE SCIENCES↗

TH-C-BRD-04: Beam Modeling and Validation with Triple and Double Gaussian Dose Kernel for Spot Scanning Proton Beams

Purpose: To present the validity of our beam modeling with double and triple Gaussian dose kernels for spot scanning proton beams in Nagoya Proton Therapy Center. This study investigates the conformance between the measurements and calculation results in absolute dose with two types of beam kernel. Methods: A dose kernel is one of the important input data required for the treatment planning software. The dose kernel is the 3D dose distribution of an infinitesimal pencil beam of protons in water and consists of integral depth doses and lateral distributions. We have adopted double and triple Gaussian model as lateral distribution in order to take account of the large angle scattering due to nuclear reaction by fitting simulated inwater lateral dose profile for needle proton beam at various depths. The fitted parameters were interpolated as a function of depth in water and were stored as a separate look-up table for the each beam energy. The process of beam modeling is based on the method of MDACC [X.R.Zhu 2013]. Results: From the comparison results between the absolute doses calculated by double Gaussian model and those measured at the center of SOBP, the difference is increased up to 3.5% in the high-energy region because the large angle scattering due to nuclear reaction is not sufficiently considered at intermediate depths in the double Gaussian model. In case of employing triple Gaussian dose kernels, the measured absolute dose at the center of SOBP agrees with calculation within ±1% regardless of the SOBP width and maximum range. Conclusion: We have demonstrated the beam modeling results of dose distribution employing double and triple Gaussian dose kernel. Treatment planning system with the triple Gaussian dose kernel has been successfully verified and applied to the patient treatment with a spot scanning technique in Nagoya Proton Therapy Center.

07 ISOTOPE AND RADIATION SOURCES↗