SU-F-J-11: Radiobiologically Optimized Patient Localization During Prostate External Beam Localization

Purpose: To present a novel positioning strategy which optimizes radiation delivery with radiobiological response knowledge, and to evaluate its application during prostate external beam radiotherapy. Methods: Ten patients with low or intermediate risk prostate cancer were evaluated retrospectively in this IRB-approved study. For each patient, a VMAT plan was generated on the planning CT (PCT) to deliver 78 Gy in 39 fractions with PTV = prostate + 7 mm margin, except for 5mm in the posterior direction. Five representative pretreatment CBCT images were selected for each patient, and prostate, rectum, and bladder were delineated on all CBCT images. Each CBCT was auto-registered to the corresponding PCT. Starting from this auto-matched position (AM-position), a search for optimal treatment position was performed utilizing a score function based on radiobiological and dosimetric indices (D98-DTV, NTCP-rectum, and NTCP-bladder) for the daily target volume (DTV), rectum, and bladder. DTV was defined as prostate + 4 mm margin to account for intra-fraction motion as well as contouring variability on CBCT. We termed the optimal treatment position the radiobiologically optimized couch shift position (ROCS-position). Results: The indices, averaged over the 10 patients’ treatment plans, were (mean±SD): 77.7±0.2 Gy (D98-PTV), 12.3±2.7% (NTCP-rectum), and 53.2±11.2% (NTCP-bladder). The corresponding values calculated on all 50 CBCT images at the AM-positions were 72.9±11.3 Gy (D98-DTV), 15.8±6.4% (NTCP-rectum), and 53.0±21.1% (NTCP-bladder), respectively. In comparison, calculated on CBCT at the ROCS-positions, the indices were 77.0±2.1 Gy (D98-DTV), 12.1±5.7% (NTCP-rectum), and 60.7±16.4% (NTCP-bladder). Compared to autoregistration, ROCS-optimization recovered dose coverage to target volume and lowered the risk to rectum. Moreover, NTCPrectum for one patient remained high after ROCS-optimization and therefore could potentially benefit from adaptive planning. Conclusion: These encouraging results illustrate the potential utility of applying radiobiologically optimized correction for online image-guided radiotherapy of prostate patients.

60 APPLIED LIFE SCIENCES↗

SU-F-J-38: Dose Rates and Preliminary Evaluation of Contouring Similarity Metrics Using 4D Cone Beam CT

Purpose: 4D imaging modalities require detailed characterization for clinical optimization. The On-Board Imager mounted on the linear accelerator was used to investigate dose rates in a tissue mimicking phantom using 4D-CBCT and assess variability of contouring similarity metrics between 4D-CT and 4D-CBCT retrospective reconstructions. Methods: A 125 kVp thoracic protocol was used. A phantom placed on a motion platform simulated a patient’s breathing cycle. An ion chamber was affixed inside the phantom’s tissue mimicking cavities (i.e. bone, lung, and soft tissue). A sinusoidal motion waveform was executed with a five second period and superior-inferior motion. Dose rates were measured at six ion chamber positions. A preliminary workflow for contouring similarity between 4D-CT and 4D-CBCT was established using a single lung SBRT patient’s historical data. Average intensity projection (Ave-IP) and maximum intensity projection (MIP) reconstructions generated offline were compared between the 4D modalities. Similarity metrics included Dice similarity coefficient (DSC), Hausdorff distance, and center of mass (COM) deviation. Two isolated lesions were evaluated in the patient’s scans: one located in the right lower lobe (ITVRLL) and one located in the left lower lobe (ITVLLL). Results: Dose rates ranged from 2.30 (lung) to 5.18 (bone) E-3 cGy/mAs. For fixed acquisition parameters, cumulative dose is inversely proportional to gantry speed. For ITVRLL, DSC were 0.70 and 0.68, Hausdorff distances were 6.11 and 5.69 mm, and COM deviations were 1.24 and 4.77 mm, for Ave-IP and MIP respectively. For ITVLLL, DSC were 0.64 and 0.75, Hausdorff distances were 10.74 and 8.00 mm, and COM deviations were 7.55 and 4.3 mm, for Ave-IP and MIP respectively. Conclusion: While the dosimetric output of 4D-CBCT is low, characterization is necessary to assure clinical optimization. A basic workflow for comparison of simulation and treatment 4D image-based contours was established. This work was partially supported by a Research Scholar Grant (RSG-15-137-01-CCE) from the American Cancer Society.

60 APPLIED LIFE SCIENCES↗

SU-G-TeP2-08: Evaluation of Plastic Scintillator Detector for Small Field Stereotactic Patient-Specific Quality Assurance

Purpose: To evaluate the performance of a commercial plastic scintillator detector (PSD) for small-field stereotactic patient-specific quality assurance using flattening-filter-free (FFF) beams. Methods: A total of ten spherical targets (volume range:[0.03cc–2cc]) were planned using Dynamic Conformal Arc(DCA-10 plans) and Volumetric Modulated Arc Therapy(VMAT-10 plans) techniques in Eclipse(AAA v.11, 1mm dose calculation grid size). Additionally, 15 previously-treated cranial and spine SRS plans were evaluated (6 DCA, 9 VMAT, volume range:[0.04cc–119.02cc]). All measurements were acquired using Varian Edge equipped with HDMLC. Three detectors were used: PinPoint ion chamber (PTW;active volume 0.015cc), Exradin W1 PSD (Standard Imaging;active volume 0.002cc), and Gafchromic EBT3 film (Ashland). PinPoint and PSD were positioned perpendicular to beam axis in a Lucy phantom (Standard Imaging). Films were placed at isocenter in solid water. Calibration films were delivered for absolute dose analysis. Results: For large spherical targets(>1.5cc) with DCA, all detectors agreed within 1% of AAA calculations. As target volume decreased, PSD measured higher doses than AAA (maximum difference: 3.3% at 0.03cc target), while PinPoint chamber measured lower doses (maximum difference:-3.8% at 0.03cc target). Inter-detector differences between pinpoint and PSD increased with decreasing target size; differences>5% were observed for targets<0.09cc. Similar trends for inter-detector behavior were observed for clinical plans. For target sizes<0.08cc, PSD measured>5% higher dose than PinPoint chamber (maximum difference: 9.25% at 0.04cc target). Film demonstrated agreement of −0.19±1.47% with PSD for all spherical targets, and agreement within −0.98±2.25% for all 15 clinical targets. Unlike DCA, VMAT plans did not show improved AAA-to-detector agreements for large targets. Conclusion: For all targets, the PSD measurements agreed with film within 1.0%, on average. For small volume targets (<0.10cc), PSD agreed with film but measured significantly higher doses (>5%) compared with the pin point ion chamber. The plastic scintillator detector appears to be suitable for accurate measurements of small SRS targets.

60 APPLIED LIFE SCIENCES↗

SU-E-J-28: Gantry Speed Significantly Affects Image Quality and Imaging Dose for 4D Cone-Beam Computed Tomography On the Varian Edge Platform

Purpose: 4D-CBCT facilitates assessment of tumor motion at treatment position. We investigated the effect of gantry speed on 4D-CBCT image quality and dose using the Varian Edge On-Board Imager (OBI). Methods: A thoracic protocol was designed using a 125 kVp spectrum. Image quality parameters were obtained via 4D acquisition using a Catphan phantom with a gating system. A sinusoidal waveform was executed with a five second period and superior-inferior motion. 4D-CBCT scans were sorted into 4 and 10 phases. Image quality metrics included spatial resolution, contrast-to-noise ratio (CNR), uniformity index (UI), Hounsfield unit (HU) sensitivity, and RMS error (RMSE) of motion amplitude. Dosimetry was accomplished using Gafchromic XR-QA2 films within a CIRS Thorax phantom. This was placed on the gating phantom using the same motion waveform. Results: High contrast resolution decreased linearly from 5.93 to 4.18 lp/cm, 6.54 to 4.18 lp/cm, and 5.19 to 3.91 lp/cm for averaged, 4 phase, and 10 phase 4DCBCT volumes respectively as gantry speed increased from 1.0 to 6.0 degs/sec. CNRs decreased linearly from 4.80 to 1.82 as the gantry speed increased from 1.0 to 6.0 degs/sec, respectively. No significant variations in UIs, HU sensitivities, or RMSEs were observed with variable gantry speed. Ion chamber measurements compared to film yielded small percent differences in plastic water regions (0.1–9.6%), larger percent differences in lung equivalent regions (7.5–34.8%), and significantly larger percent differences in bone equivalent regions (119.1–137.3%). Ion chamber measurements decreased from 17.29 to 2.89 cGy with increasing gantry speed from 1.0 to 6.0 degs/sec. Conclusion: Maintaining technique factors while changing gantry speed changes the number of projections used for reconstruction. Increasing the number of projections by decreasing gantry speed decreases noise, however, dose is increased. The future of 4DCBCT’s clinical utility relies on further investigation of image optimization.

60 APPLIED LIFE SCIENCES↗

SU-E-T-321: Dosimetric Evaluation of Non-Coplanar Arcs in VMAT Planning for SBRT Lung Cases: A Planning Study

Purpose: Investigate use of standardized non-coplanar arcs to improve plan quality in lung Stereotactic Body Radiation Therapy(SBRT) VMAT planning. Methods: VMAT planning was performed for 9 patients previously treated with SBRT for peripheral lung tumors (tumor size:12.7cc to 32.5cc). For each patient, 7 VMAT plans (couch rotation values:0,5,10,15,20,25,and 30 deg) were generated; the coplanar plans were pushed to meet the RTOG0915 constraints and each non-coplanar plans utilized the same optimization constraints. The following plan dose metrics were used (taken from RTOG 0915): D-2cm: the maximum dose at 2 cm from the PTV, conformality index (CI), gradient index (GI), lung volume receiving 5 Gy (V5) and lung volume receiving 20 Gy (V20). The couch collision clearance was checked for each plan through a dry run using the couch position from the patient’s treatment. Results: Of the 9 cases, one coplanar plan failed to meet two protocol guidelines (both gradient index and D-2cm parameter), and an additional plan failed the D-2cm parameter. When introducing at least 5 degree couch rotation, all plans met the protocol guidelines. The largest feasible couch angle available was 15 to 20 degrees due to gantry collision issues. Non-coplanar plans resulted in the average (standard deviation) reduction of the following metrics: GI by 7.3% (3.7%); lung V20 by 11.1% (3.2%); D-2cm by 12.7% (3.9%). The CI was unchanged (−0.3%±0.6%), and lung V5 increased (3.8%±8.2%). Conclusion: The use of couch rotations as little as 5 degrees allows for plan quality that will meet RTOG0915 constraints while reducing D-2cm, GI, and lung V20. Using default couch rotations while planning SBRT cases will allow for more efficient planning with the stated goal of meeting RTOG0915 criteria for all clinical cases. Gantry clearance checks in the treatment room may be necessary to ensure safe treatments for larger couch rotation values.

60 APPLIED LIFE SCIENCES↗

QCD and Strongly Coupled Gauge Theories: Challenges and Perspectives

We highlight the progress, current status, and open challenges of QCD-driven physics, in theory and in experiment. We discuss how the strong interaction is intimately connected to a broad sweep of physical problems, in settings ranging from astrophysics and cosmology to stongly-coupled, complex systems in particle and condensed-matter physics, as well as to searches for physics beyond the Standard Model. We also discuss how success in describing the strong interaction impacts other fields, and, in turn, how such subjects can impact studies of the strong interaction. In the course of the work we offer a perspective on the many research streams which flow into and out of QCD, as well as a vision for future developments.

72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS↗

SU-F-BRE-10: Methods to Simulate and Measure the Attenuation for Modeling a Couch Top with Rails for FFF Treatment Delivery On the Varian Edge Linac

Purpose: To measure attenuation for modelling of the KVue Couchtop for 6X and 10X FFF SRS/SBRT treatment Methods: Treatment planning simulation studies were done using 6X FFF beams to estimate the dosimetric impact of KVue couchtops (including the Q-Fix IGRT [carbon fiber] and Calypso [nonconductive Kevlar material]) with a structure model obtained from a research workstation (Eclipse, advanced planning interface (API) v13). Prior to installation on the Varian Edge linac, the couchtop along with (Kevlar) rails were CT scanned with the rails at various positions. An additional scan with the couchtop 15cm above the CT table top was obtained with 20cm solid water to facilitate precised/indexed data acquisition. Measurements for attenuation were obtained for field sizes of 2, 4 and 10 cm{sup 2} at 42 gantry angles including 6 pairs of opposing fields and other angles for oblique delivery where the beams traversed the couchtop and or rails. The delivery was fully automated with xml scripts running in developer mode. The results were then used to determine an accurate structure model for AAA (Eclipse v11) planning of IMRT and RapidArc delivery. Results: The planning simulation relative dose attenuation for oblique entry was not significantly different than the Exact IGRT or BrainLab iBeam couch except that the rails added 6% additional attenuation. The relative attenuation measurements for PA, PA (rails: inner position), oblique, oblique (rails: outer position), oblique (rails: inner position) were: −2.0%, −2.5%, −15.6%, −2.5%, −5.0% for 6X FFF and −1.4%, −1.5%, −12.2%, − 2.5%, −5.0% for 10X FFF with slight decrease in attenuation versus field size. A Couch structure model (with HU values) was developed. Calculation compared to measurement showed good agreement except for oblique (rails: outer position) where differences approached a magnitude of 6%. Conclusion: A model of the couch structures has been developed accounting for attenuation for FFF beams.

07 ISOTOPE AND RADIATION SOURCES↗

SU-E-J-152: Evaluation of TrueBeam OBI V. 1.5 CBCT Performance in An Adaptive RT Environment

Purpose: To evaluate the image quality and imaging dose of the Varian TrueBeam OBIv.1.5 CBCT system in a clinical adaptive radiation therapy environment, simulated by changing phantom thickness. Methods: Various OBI CBCT protocols(Head, Pelvis, Thorax, Spotlight) were used to acquire images of Catphan504 phantom(nominal phantom thickness and 10 cm additional phantom thickness). The images were analyzed for low contrast detectability(CNR), uniformity(UI), and HU sensitivity. These results were compared to the same image sets for planning CT(pCT)(GE LightSpeed 16- slice). Imaging dose measurements were performed with Gafchromic XRQA2 film for various OBI protocols (Pelvis, Thorax, Spotlight) in a pelvic-sized phantom(nominal thickness and 4cm additional thickness). Dose measurements were acquired in the interior and at the surface of the phantom. Results: The nominal CNR[additional thickness CNR] for OBI was—Pelvis:1.45[0.81],Thorax:0.86[0.48], Spotlight:0.67[0.39],Head:0.28 [0.10]. The nominal CNR[additional thickness CNR] for pCT was— Pelvis:0.87[0.41],Head:0.60[0.22]. The nominal UI[additional thickness UI] for OBI was—Pelvis:11.5[24.1],Thorax:17.0[20.6], Spotlight:23.2[23.2], Head:15.6[59.9]. The nominal UI[additional thickness UI] for pCT was— Pelvis:9.2[8.6],Head:2.1[2.9]. The HU difference(averaged over all material inserts) between nominal and additional thickness scans for OBI: 8.26HU(Pelvis), 33.39HU(Thorax), 178.98HU(Head), 108.20HU (Spotlight); for pCT: 16.00HU(Pelvis), 19.85HU(Head). Uncertainties in electron density were calculated based on HU values with varying phantom thickness. Average electron-density deviations (ρ(water)=1)for GE-Pelvis, GE-Head, OBI-Pelvis, OBI-Thorax, OBI-Spotlight, and OBI-Head were: 0.0182, 0.0180, 0.0058, 0.0478, 0.2750, and 0.3115, respectively.The average phantom interior dose was(OBI-nominal):2.35cGy(Pelvis), 0.60cGy(Thorax), 1.87cGy(Spotlight); OBI-increased thickness: 1.77cGy(Pelvis), 0.43cGy(Thorax), 1.53cGy (Spotlight). Average surface dose(OBI-nominal): 2.29cGy(Pelvis), 0.56cGy(Thorax), 1.79cGy (Spotlight); OBI-increased thickness: 1.94cGy(Pelvis), 0.48cGy(Thorax), 1.47cGy (Spotlight). Conclusion: The OBI-Pelvis protocol offered comparable CNR and HU constancy to pCT for each geometry; other protocols, particularly Spotlight and Head, exhibited lower HU constancy and CNR. The uniformity of pCT was superior to OBI for all protocols. CNR and UI were degraded for both systems/scan types with increased thickness. The OBI interior dose decreased by approximately 30% with additional thickness. This work was funded, in part, under a grant with the Pennsylvania Department of Health. The Department of Health specifically declaims responsibility for any analyses, interpretations, or conclusions.

60 APPLIED LIFE SCIENCES↗

SU-E-T-406: Use of TrueBeam Developer Mode and API to Increase the Efficiency and Accuracy of Commissioning Measurements for the Varian EDGE Stereotactic Linac

Purpose: The Varian EDGE machine is a new stereotactic platform, combining Calypso and VisionRT localization systems with a stereotactic linac. The system includes TrueBeam DeveloperMode, making possible the use of XML-scripting for automation of linac-related tasks. This study details the use of DeveloperMode to automate commissioning tasks for Varian EDGE, thereby improving efficiency and measurement consistency. Methods: XML-scripting was used for various commissioning tasks,including couch model verification,beam-scanning,and isocenter verification. For couch measurements, point measurements were acquired for several field sizes (2×2,4×4,10×10cm{sup 2}) at 42 gantry angles for two couch-models. Measurements were acquired with variations in couch position(rails in/out,couch shifted in each of motion axes) compared to treatment planning system(TPS)-calculated values,which were logged automatically through advanced planning interface(API) scripting functionality. For beam scanning, XML-scripts were used to create custom MLC-apertures. For isocenter verification, XML-scripts were used to automate various Winston-Lutz-type tests. Results: For couch measurements, the time required for each set of angles was approximately 9 minutes. Without scripting, each set required approximately 12 minutes. Automated measurements required only one physicist, while manual measurements required at least two physicists to handle linac positions/beams and data recording. MLC apertures were generated outside of the TPS,and with the .xml file format, double-checking without use of TPS/operator console was possible. Similar time efficiency gains were found for isocenter verification measurements Conclusion: The use of XML scripting in TrueBeam DeveloperMode allows for efficient and accurate data acquisition during commissioning. The efficiency improvement is most pronounced for iterative measurements, exemplified by the time savings for couch modeling measurements(approximately 10 hours). The scripting also allowed for creation of the files in advance without requiring access to TPS. The API scripting functionality enabled efficient creation/mining of TPS data. Finally, automation reduces the potential for human error in entering linac values at the machine console,and the script provides a log of measurements acquired for each session. This research was supported in part by a grant from Varian Medical Systems, Palo Alto, CA.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Design of Genomic Signatures of Pathogen Identification & Characterization

This chapter will address some of the many issues associated with the identification of signatures based on genomic DNA/RNA, which can be used to identify and characterize pathogens for biodefense and microbial forensic goals. For the purposes of this chapter, we define a signature as one or more strings of contiguous genomic DNA or RNA bases that are sufficient to identify a pathogenic target of interest at the desired resolution and which could be instantiated with particular detection chemistry on a particular platform. The target may be a whole organism, an individual functional mechanism (e.g., a toxin gene), or simply a nucleic acid indicative of the organism. The desired resolution will vary with each program's goals but could easily range from family to genus to species to strain to isolate. The resolution may not be taxonomically based but rather pan-mechanistic in nature: detecting virulence or antibiotic-resistance genes shared by multiple microbes. Entire industries exist around different detection chemistries and instrument platforms for identification of pathogens, and we will only briefly mention a few of the techniques that we have used at Lawrence Livermore National Laboratory (LLNL) to support our biosecurity-related work since 2000. Most nucleic acid based detection chemistries involve the ability to isolate and amplify the signature target region(s), combined with a technique to detect the amplification. Genomic signature based identification techniques have the advantage of being precise, highly sensitive and relatively fast in comparison to biochemical typing methods and protein signatures. Classical biochemical typing methods were developed long before knowledge of DNA and resulted in dozens of tests (Gram's stain, differential growth characteristics media, etc.) that could be used to roughly characterize the major known pathogens (of course some are uncultivable). These tests could take many days to complete and precise resolution of species and strains is not always possible. In contrast, protein recognition signatures composed of antibodies or synthetic high-affinity ligands offer extremely fast results but require a large quantity of the target to be present. False positives/negatives are also a problem with some protein-based techniques (home pregnancy kits use this basic approach).

59 BASIC BIOLOGICAL SCIENCES↗

TMTI Task 1.6 Genetic Engineering Methods and Detection

A large number of GE techniques can be adapted from other microorganisms to biothreat bacteria and viruses. Detection of GE in a microorganism increases in difficulty as the size of the genetic change decreases. In addition to the size of the engineered change, the consensus genomic sequence of the microorganism can impact the difficulty of detecting an engineered change in genomes that are highly variable from strain to strain. This problem will require comprehensive databases of whole genome sequences for more genetically variable biothreat bacteria and viruses. Preliminary work with microarrays for detecting synthetic elements or virulence genes and analytic bioinformatic approaches for whole genome sequence comparison to detect genetic engineering show promise for attacking this difficult problem but a large amount of future work remains.

59 BASIC BIOLOGICAL SCIENCES↗