Haque, Razi

Publications and source records attributed to Haque, Razi.

Swab Testing Results

1. All 3D printed swabs bend to at least 90° without breaking. 2. All 3D printed swabs are at least as strong under tensile testing as their commercial counterparts. a. The Abiogenix FAST Spiral NP and EnvisionTEC swabs retain their tensile strength after sterilization and after 7 days of aging. b. For the Formlabs swabs, steam sterilization significantly reduces their tensile strength, which mostly recovers after 7 days of aging. However, the swabs are stronger than commercial ones at all times. c. The Resolution Medical Lattice swabs retain their tensile strength after 5 days of aging. Five different swab types were tested: 1. Abiogenix FAST Spiral NP Swab; 2. Formlabs NP Swab; 3. EnvisionTEC NP Swab; 4. Resolution Medical Lattice Swab; and 5. Fosun Pharma NP Swab The Abiogenix FAST Spiral NP, EnvisionTEC NP, Resolution Medical Lattice, and Fosun Pharma NP swabs were provided by their respective suppliers. The Formlabs NP swabs were printed at LLNL using the Form 3B Printer and Formlabs Surgical Guide resin

59 BASIC BIOLOGICAL SCIENCES↗

Swab Testing Results Summary

Swabs Tested: 1. FAST Spiral NP Swab, designed and supplied by Abiogenix, printed from polyamide 11 (Nylon) using HP Multi Jet Fusion printers. 2. NP Swab designed by USF Health and Northwell Health in collaboration with Formlabs. Printed at LLNL using the Form 3B stereolithography printer in Formlabs Surgical Guide resin.

42 ENGINEERING↗

Swab Tensile Testing Results and Procedures - PA11 Rev. 10 Swabs

This tensile test is to mimic the worst-case scenario of the swab being caught on an obstruction when being pulled out of the nasopharyngeal space. It is to determine how much tensile force the swab can withstand without breaking. All swabs were provided by HP and were the PA11 material and Rev 10 design. All of the swabs were tested using the protocol outlined in Section 2 (Tensile Testing Procedure). Not all of the swabs were pre-treated using the Autoclave. For those swabs that were pre-treated using the Autoclave, the tensile tests were performed within 6 hours of completion of the Autoclave pre-treatment, with the exception of the “Aged” swabs. For the “Aged” swabs, the tensile tests were performed 7 days after the Autoclave pre -treatment. The “Aged” swabs that were not pre-treated using the Autoclave, were tested 7 days after those that were not “Aged.”

36 MATERIALS SCIENCE↗

Swab Tensile Testing Results and Procedures

This tensile test is to mimic the worst-case scenario of the swab being caught on an obstruction when being pulled out of the nasopharyngeal space. It is to determine how much tensile force the swab can withstand without breaking. A variety of different swabs (different materials and different geometries) were tested using the protocol outlined in Section 2 (Tensile Testing Procedure). Not all of the swabs were pre-treated using the Autoclave. For those swabs that were pre-treated using the Autoclave, the tensile tests were performed within 6 hours of completion of the Autoclave pre-treatment, with the exception of the “Aged” swabs. For the “Aged” swabs, the tensile tests were performed 7 days after the Autoclave pre -treatment. All swabs were provided by HP.

36 MATERIALS SCIENCE↗

Swab Tensile Testing - HP Rev 10 Summary

This tensile test is to mimic the worst-case scenario of the swab being caught on an obstruction when being pulled out of the nasopharyngeal space. It is to determine how much tensile force the swab can withstand without breaking.

42 ENGINEERING↗

Swab Testing Results- HP PA11 Rev 8 Summary

This tensile test is to mimic the worst-case scenario of the swab being caught on an obstruction when being pulled out of the nasopharyngeal space. It is to determine how much tensile force the swab can withstand without breaking.

42 ENGINEERING↗

Swab Tensile Testing and Procedures

This tensile test is to mimic the worst-case scenario of the swab being caught on an obstruction when being pulled out of the nasopharyngeal space. It is to determine how much tensile force the swab can withstand without breaking. A variety of different swabs (different materials and different geometries) were tested using the protocol outlined in Section 2 (Tensile Testing Procedure). Not all of the swabs were pre-treated using the Autoclave. For those swabs that were pre-treated using the Autoclave, the tensile tests were performed within 6 hours of completion of the Autoclave pre-treatment, unless otherwise noted. The FormLabs-USF and FormLabs-Northwell were printed at LLNL using FormLabs Surgical Grade V1 Resin on a Form 3B Printer. Results Summary: All FormLabs design and treatment variations (e.g. autoclaving and “aging”) had higher tensile strength compared commercial swabs. Autoclaving degraded the tensile strength. “Aging” after autoclaving improved the tensile strength. “Aging” without autoclaving degraded the tensile strength. No differences were observed in tensile strength with different autoclaving protocols. No significant differences were observed in tensile strength for different batches.

59 BASIC BIOLOGICAL SCIENCES↗

Quantitative form and fit of N95 filtering facepiece respirators are retained after dry and humid heat treatments for coronavirus deactivation

Re-use of filtering facepiece respirators (FFRs, commonly referred to as N95s) normally meant for single use only is becoming common in healthcare facilities due to shortages caused by the COVID19 pandemic. Here we report that mouse hepatitis virus (MHV) initially seeded on FFR filter material is inactivated (6 log reduction as measured by 50% tissue culture infective dose (TCID50)) after dry heating at 75 °C for 30 minutes. We also find that the quantitative fit of FFRs after heat treatment at this temperature, under dry conditions or at 90% relative humidity, is not affected by single or ten heating cycles. Previous studies have reported that the filtration efficiency of FFR filters is not negatively impacted by these heating conditions. These results suggest that thermal inactivation of coronaviruses is a potentially rapid and widely deployable method to re-use N95 FFRs in emergency situations where reusing FFRs is a necessity and broad-spectrum sterilization is unavailable. However, we also found that a heat source that emits radiation (e.g., an exposed heating element) results in rapid qualitative degradation of the FFR. Finally, we discuss differences in the results reported here and other recent studies investing heat as a means to recycle FFRs and suggest that overall wear time and donning/doffing cycles are important factors that need to be considered.

36 MATERIALS SCIENCE↗

Swab Tensile Testing Results [Slides]

HP PA11 material swabs predominantly broke at the joint between the head and the neck, although there was significant variation in the break locations. HP PA12, Injection Molded, and FormLabs (Surgical Guide V1 Resin) material swabs predominately broke in the neck region.

36 MATERIALS SCIENCE↗

Swab Tensile Testing Results and Procedures

This tensile test is to mimic the worst-case scenario of the swab being caught on an obstruction when being pulled out of the nasopharyngeal space. It is to determine how much tensile force the swab can withstand without breaking. A variety of different swabs (different materials and different geometries) were tested using the protocol outlined in Section 2 (Tensile Testing Procedure). Not all of the swabs were pre-treated using the Autoclave. For those swabs that were pre-treated using the Autoclave, the tensile tests were performed within 6 hours of completion of the Autoclave pre-treatment, with the exception of the Injection Molded samples. For the Injection Molded samples, the tensile tests were performed within 24 hours of the completion of the Autoclave pre-treatment. The FormLabs-USF and FormLabs-Northwell were printed at LLNL using FormLabs Surgical Grade V1 Resin on a Form 3B Printer. All other samples were provided by HP.

36 MATERIALS SCIENCE↗

A microfabricated, 3D-sharpened silicon shuttle for insertion of flexible electrode arrays through dura mater into brain

Electrode arrays for chronic implantation in the brain are a critical technology in both neuroscience and medicine. Recently, flexible, thin-film polymer electrode arrays have shown promise in facilitating stable, single-unit recordings spanning months in rats. While array flexibility enhances integration with neural tissue, it also requires removal of the dura mater, the tough membrane surrounding the brain, and temporary bracing to penetrate the brain parenchyma. Durotomy increases brain swelling, vascular damage, and surgical time. Insertion using a bracing shuttle results in additional vascular damage and brain compression, which increase with device diameter; while a higher-diameter shuttle will have a higher critical load and more likely penetrate dura, it will damage more brain parenchyma and vasculature. One way to penetrate the intact dura and limit tissue compression without increasing shuttle diameter is to reduce the force required for insertion by sharpening the shuttle tip. We describe a novel design and fabrication process to create silicon insertion shuttles that are sharp in three dimensions and can penetrate rat dura, for faster, easier, and less damaging implantation of polymer arrays. Sharpened profiles are obtained by reflowing patterned photoresist, then transferring its sloped profile to silicon with dry etches. We demonstrate that sharpened shuttles can reliably implant polymer probes through dura to yield high quality single unit and local field potential recordings for at least 95 days. On insertion directly through dura, tissue compression is minimal. Significance. This is the first demonstration of a rat dural-penetrating array for chronic recording. This device obviates the need for a durotomy, reducing surgical time and risk of damage to the blood-brain barrier. This is an improvement to state-of-the-art flexible polymer electrode arrays that facilitates their implantation, particularly in multi-site recording experiments. This sharpening process can also be integrated into silicon electrode array fabrication.

60 APPLIED LIFE SCIENCES↗