Radiometric calibration of SPOT 2 HRV - A comparison of three methods

Three methods for determining an absolute radiometric calibration of a spacecraft optical sensor are compared. They are the well-known reflectance-based and radiance-based methods and a new method based on measurements of the ratio of diffuse-to-global irradiance at the ground. The latter will be described in detail and the comparison of the three approaches will be made with reference to the SPOT-2 HRV cameras for a field campaign 1990-06-19 through 1990-06-24 at the White Sands Missile Range in New Mexico.

Biggar, Stuart F.

Results of calibrations of the NOAA-11 AVHRR made by reference to calibrated SPOT imagery at White Sands, N.M

The calibration method reported here makes use of the reflectances of several large, uniform areas determined from calibrated and atmospherically corrected SPOT Haute Resolution Visible (HRV) scenes of White Sands, New Mexico. These reflectances were used to predict the radiances in the first two channels of the NOAA-11 Advanced Very High Resolution Radiometer (AVHRR). The digital counts in the AVHRR image corresponding to these known reflectance areas were determined by the use of two image registration techniques. The plots of digital counts versus pixel radiance provided the calibration gains and offsets for the AVHRR. A reduction in the gains of 4 and 13 percent in channels 1 and 2 respectively was found during the period 1988-11-19 to 1990-6-21. An error budget is presented for the method and is extended to the case of cross-calibrating sensors on the same orbital platform in the Earth Observing System (EOS) era.

Nianzeng, Che

Radiometric calibration of an airborne multispectral scanner

The absolute radiometric calibration of the NS001 Thematic Mapper Simulator reflective channels was examined based on laboratory tests and in-flight comparisons to ground measurements. The NS001 data are calibrated in-flight by reference to the NS001 internal integrating sphere source. This source's power supply or monitoring circuitry exhibited greater instability in-flight during 1988-1989 than in the laboratory. Extrapolating laboratory behavior to in-flight data resulted in 7-20 percent radiance errors relative to ground measurements and atmospheric modeling. Assuming constancy in the source's output between laboraotry and in-flight resulted in generally smaller errors. Upgrades to the source's power supply and monitoring circuitry in 1990 improved its in-flight stability, though in-flight ground reflectance based calibration tests have not yet been performed.

Markham, Brian L.

Mapping surface energy balance components by combining Landsat Thematic Mapper and ground-based meteorological data

Surface energy balance components were evaluated by combining satellite-based spectral data with on-site measurements of solar irradiance, air temperature, wind speed, and vapor pressure. Maps of latent heat flux density and net radiant flux density were produced using Landsat TM data for three dates. The TM-based estimates differed from Bowen-ratio and aircraft-based estimates by less than 12 percent over mature fields of cotton, wheat, and alfalfa.

Moran, M. Susan

Surface reflectance factor retrieval from Thematic Mapper data

Based on the absolute radiometric calibration of the TM and the use of a radiative transfer program for atmospheric correction, ground reflectances were retrieved for several fields of crops and bare soil in TM bands 1-4 for six TM scenes acquired over a 12-month period. These reflectances were compared to those measured using ground-based and low-altitude, aircraft-mounted radiometers. When, for four TM acquisitions, the comparison was made between areas that had been carefully selected for their high uniformity, the reflectance factors agreed to + or - 0.01 over the reflectance range 0.02-0.55. When the comparison was made for two of the above acquisitions and two others on different dates, for larger areas not carefully selected to be of uniform reflectance, the reflectance factors agreed to + or - 0.02 (1 sigma RMS), over same reflectance range.

Holm, Ronald G.

Field calibration of reference reflectance panels

A procedure for calibrating reference reflectance panels using the sun as the radiation source and a pressed-polytetrafluoroethylene powder standard is described. The directional/directional reflectance factor and the directional/hemispheric reflectance factor are examined. Directional/directional voltage responses for pressed-halon are analyzed. Three painted BaSO4 and one painted halon were calibrated using the proposed procedure. The effects of diffuse irradiance on reflectance-factor measurements are investigated. It is determined that the method has an accuracy on the order of 1 percent. The advantages and disadvantages of this method are discussed.

Jackson, Ray D.