CCN comparisons of static diffusion chambers

Nine CCN counters of the static diffusion (SDC) type were compared with one another and with continuous flow diffusion (CFD) chambers. The nine SDCs showed a considerable amount of variation, largely attributable to newness and/or lack of prior calibration of some units. The five more consistent instruments agreed quite well, to within at least 20 percent of the NRL mobility analyzer and to within 10 percent at 1 percent supersaturation. There was satisfactory agreement between the more reliable SDC and CFD chambers.

Jiusto, J. E.

CCN-supersaturation spectra slopes (k)

Theoretically the slope k of a CCN-supesaturation spectrum should equal two thirds of the slope of the total (soluble) aerosol size distribution. Workshop results tended to verify this relation. The k values are markedly different depending on whether one is measuring ambient CCN concentrations at supersaturations S above or below approximately 0.1-0.2%. The larger k values for S approximately 0.1% is consistent with the greater decrease in large particle concentration with increasing size. It is concluded that over the S range of 0.02% to 2%, two power fits (and k values) may sometimes suffice for a reasonable approximation of the CCN distribution. At other times, and with laboratory generated aeosols, such an approach is inadequate and requires refinement.

Jiusto, J. E.