Arcsecond grating drive mechanism for operation at 4 K
A grating drive mechanism that achieves arcsecond positional control at both room temperature and 4 K is described. The mechanism accommodates eight equally spaced facets on a 360-deg drum, with each facet having arcsecond control over an angular range of +/-3 deg. The cryogenic portion consists of dry lubricated angular contact bearings, a dc brushless torque motor, and a compact magnetic position sensor capable of less than 1 arcsec rms resolution. Test results demonstrate angular repeatability of less than 2 arcsec rms at 4 K and static power dissipation to the helium bath of less than 100 microW.
Downey, C. H.↗