In situ X-ray and IR probes relevant to Earth science at the Advanced Light Source at Lawrence Berkeley Laboratory

Access to synchrotron X-ray facilities has become an important aspect for many disciplines in experimental Earth science. This is especially important for studies that rely on probing samples in situ under natural conditions different from the ones found at the surface of the Earth. The non-ambient condition Earth science program at the Advanced Light Source (ALS), Lawrence Berkeley National Laboratory, offers a variety of tools utilizing the infra-red and hard X-ray spectrum that allow Earth scientists to probe Earth and environmental materials at variable conditions of pressure, stress, temperature, atmospheric composition, and humidity. These facilities are important tools for the user community in that they offer not only considerable capacity (non-ambient condition diffraction) but also complementary (IR spectroscopy, microtomography), and in some cases unique (Laue microdiffraction) instruments. The availability of the ALS’ in situ probes to the Earth science community grows especially critical during the ongoing dark time of the Advanced Photon Source in Chicago, which massively reduces available in situ synchrotron user time in North America.

58 GEOSCIENCES↗

The hard x-ray nanotomography microscope at the advanced light source

Beamline 11.3.1 at the Advanced Light Source is a tender/hard (6–17 keV) x-ray bend magnet beamline recently re-purposed with a new full-field, nanoscale transmission x-ray microscope. The microscope is designed to image composite and porous materials possessing a submicrometer structure and compositional heterogeneity that determine materials’ performance and geologic behavior. The theoretical and achieved resolutions are 55 and <100 nm, respectively. The microscope is used in tandem with a <25 nm eccentricity rotation stage for high-resolution volume imaging using nanoscale computed tomography. The system also features a novel bipolar illumination condenser for the illumination of an ~100 μm spot of interest on the sample, followed by a phase-type zone plate magnifying objective of ~52 µm field of view and a phase detection ring. The zone plate serves as the system objective and magnifies the sample with projection onto an indirect x-ray detection system, consisting of a polished single crystal CsI(Tl) scintillator and a range of high-quality Plan Fluorite visible light objectives. The objectives project the final visible light image onto a water-cooled CMOS 2048 × 2048-pixel 2 detector. Here, in this paper, we will discuss the salient features of this instrument and describe early results from imaging the internal three-dimensional microstructure and nanostructure of target materials, including fiber-reinforced composites and geomaterials.

47 OTHER INSTRUMENTATION↗

X-ray Micro-Tomography of Ablative Heat Shield Materials

X-ray micro-tomography is a non-destructive characterization technique that allows imaging of materials structures with voxel sizes in the micrometer range. This level of resolution makes the technique very attractive for imaging porous ablators used in hypersonic entry systems. Besides providing a high fidelity description of the material architecture, micro-tomography enables computations of bulk material properties and simulations of micro-scale phenomena. This presentation provides an overview of a collaborative effort between NASA Ames Research Center and Lawrence Berkeley National Laboratory, aimed at developing micro-tomography experiments and simulations for porous ablative materials. Measurements are carried using x-rays from the Advanced Light Source at Berkeley Lab on different classes of ablative materials used in NASA entry systems. Challenges, strengths and limitations of the technique for imaging materials such as lightweight carbon-phenolic systems and woven textiles are discussed. Computational tools developed to perform numerical simulations based on micro-tomography are described. These enable computations of material properties such as permeability, thermal and radiative conductivity, tortuosity and other parameters that are used in ablator response models. Finally, we present the design of environmental cells that enable imaging materials under simulated operational conditions, such as high temperature, mechanical loads and oxidizing atmospheres.Keywords: Micro-tomography, Porous media, Ablation

tomography↗