Design and Characterization of a Sodium Iodide (NaI) Array for Coincidence Measurements
An array of sodium iodide detectors (NaI) was constructed to record time-correlated gamma ray measurements. The NaI Array consists of twelve 2×4×16-inch detectors. When the array is placed close to an inspected object, this arrangement provides high efficiency for gamma ray detection, particularly for high-energy gamma rays. Data are collected in time-tagged, list-mode format, which is suitable for the application of a variety of coincidence analysis methods. Automated energy calibration of each detector is accomplished by utilizing gamma radiation from thoriated welding rods that are embedded in the system. The energy resolution of spectra that are produced by the NaI array is 7.3% at 662 keV. The data are processed to sum gamma-ray energies for simultaneous detections in adjacent detectors, which are dominated by the escape of scattered photons from one detector and subsequent absorption in an adjacent detector. When data are processed in this way, the photopeak efficiency exceeds 20% over the range 400 keV to 10 MeV; the total efficiency, which includes the Compton continuum and single and double escape peaks, exceeds 70% over this energy range. Utilization of time correlations enables isolation of spectral features that are associated with the same decay cascade or fission event. Characterization of the response of the NaI Array to both gamma rays and neutrons with the Gamma Detector Response and Analysis Software (GADRAS) enables accurate computation for virtually any radiation sources.