Search NASA⌕ Search

Engineering topics

Mozhayev, Andrey V.

Publications and source records attributed to Mozhayev, Andrey V..

Development and testing of a fission-like neutron field for dosimetry and instrument calibrations based on deuterium-tritium generator

For nearly five decades 252 Cf sources have been used for testing and calibration over a wide range of neutron detection devices used for nuclear safety and radiation protection. However, the vastly increased cost of 252 Cf sources, its short half-life, and concerns about future shortage have prompted nuclear facilities and radiological calibration laboratories that rely on high-intensity neutron sources to seek alternatives. There appear to be no other radionuclide neutron sources that would match “like-to-like” 252 Cf performance characteristics. Alternatives of producing a fission-like neutron spectrum based on commercially available deuterium-tritium (D-T) generators, and without any fissionable materials or beryllium, have been explored. This paper describes efforts of designing the spectrum shaping assembly, its construction and installation. The produced neutron field was evaluated both from the perspective of the neutron fluence spectrum and the resulting dosimetric outcome. Finally, it was tested for practical application by examining response of various health physics instruments and personal monitoring devices in comparison to the response of those devices when exposed to 252 Cf. In conclusion, the response of survey-type instruments tested within the prototype, surrogate fission, reference field are acceptably close to their response in the targeted 252 Cf field.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

A Fission-like Neutron Spectrum Shaping Assembly based on D-T Generator

For nearly five decades, 252 Cf sources have been used for testing and calibration over a wide range of neutron detection devices used for nuclear safety and radiation protection. However, the vastly increased cost of 252 Cf sources, its short half-life, and concerns about future shortage have prompted nuclear facilities and radiological calibration laboratories that rely on high-intensity neutron sources to seek alternatives. While there appear to be no other radionuclide neutron sources that would match the 252 Cf performance characteristics, all of them, including 252 Cf, present similar concerns in terms of hazards, safety, and security. As a possible solution to this issue, the Pacific Northwest National Laboratory (PNNL) developed a low-risk, cost-efficient alternative to produce fission-like spectra based on commercially available deuterium-tritium (D-T) neutron generators. The alternative of producing a fission-like neutron spectrum without any fissionable materials or beryllium would enable the nuclear facilities across the entire complex to eliminate their reliance on high intensity 252 Cf sources for nuclear safety and radiation protection applications, substantially reducing risks and associated costs. This report describes efforts of designing the spectrum shaping assembly, its construction and characterization of the neutron field produced. Performance of this newly developed D-T generator based neutron field is demonstrated via calibration of a criticality detector, neutron survey instruments, and personal dosimeters verified against current standardized fields produced by 252 Cf spontaneous fission sources.

42 ENGINEERING↗

Configuration and impurity quantification of AmLi sources using radiography and gamma spectroscopy

This paper presents the non-destructive assay of two AmLi sources used for detector characterization and active interrogation at PNNL, identified as MRC-101 and MRC-103. First, a detailed description of the internal configuration of 2724-BT encapsulated MRC sources is established. X-ray radiographs of the AmLi sources are reported for the first time to verify the dimensions and orientation of the encapsulation. Notably, a density variation was observed as a contrast change across the height of the source cavities. Contrary to commonly made assumptions in literature, it is indirectly observed that the AmLi source/target material only occupies a portion of the source cavity. The intensity of the prompt gamma-rays resulting from the inelastic scatter of alpha particles on 7Li was used to calculate the mass ratios of LiH:AmO2. The gamma-ray spectra revealed the presence of numerous impurities in both source and target material (243Am, 237Np, 154Eu, 23Na, and 9Be). Neutron-gamma production ratios were applied to the intensity of the Doppler-broadened peaks from alpha-induced reactions with 7Li, 9Be, and 23Na to calculate neutron rates for both sources. A neutron/alpha production yield (derived in this work) was then related to the atomic concentration of these isotopes to find the approximate masses within each source. Ultimately, the AmLi source density was estimated with prompt gamma analysis based on partial source volumes as 0.45±0.08 g/cm^3.

DTRA, AmLi, neutron↗

Characterization of Two Extraordinary AmLi Neutron Sources

Two legacy AmLi sources, MRC-101 and MRC-103, are routinely used for neutron detector characterization and testing campaigns that Pacific Northwest National Laboratory manages for the Defense Threat Reduction Agency within the U.S. Department of Defense. However, applicability of source measurement data has been substantially limited due to the lack of original manufacturer documentation and cross comparison to other sources. This paper summarizes results of the measurement campaign conducted to better quantify source radionuclide content, neutron energy spectrum and emission rate of the AmLi sources. The source characterization study consisted of three major consecutive measurement exercises. First, gamma spectrometry measurements using a high purity germanium detector were used to estimate the radionuclide content and potential impurities of the AmLi sources. Then neutron spectral measurements were taken to determine the unique neutron energy distribution. Finally, assessments of the total neutron yields of the sources were made with measurements in a neutron well counter. The neutron detection efficiency for these measurements was determined via detailed Monte Carlo modeling based on the neutron energy distribution measured with the rotating neutron spectrometer. The most exciting and unexpected finding of the study was a significant contamination of the source material with beryllium, contributing up to 17% to the total neutron emission rate and correspondingly changing the neutron energy spectrum. Also, the obtained lithium-only portion of the measured energy distribution of the sources was compared against several AmLi spectra previously reported by different researchers.

AWCC, AmLi, beryllium contamination, neutron spect↗