Radon-222 Charcoal Canister Steady State Model Calibrations Performed in a Highly Controlled Environmental Chamber and a Natural Indoor Environment

Charcoal canisters are a common method of 222 Rn screening. The calibrations of different batches of activated charcoal used in different canister designs are typically performed in large volume controlled environmental chambers with known and controlled radon concentration, temperature, and humidity. Radon screening could be facilitated in locations without ready access to environmental chambers if canisters could be accurately calibrated and undergo quality control in less controlled environments. This study compares charcoal canister calibrations from a highly controlled radon chamber with calibrations from a basement storage area experiencing temporally varying radon. In addition, the impacts of exposure time and spectral region of interest selection on calibration accuracy are investigated by comparing calibrations calculated using three different choices. Approximately 30 mo after calibrations were completed, groups of canisters were exposed for different durations of time to investigate measurement accuracy and calibration validity over an extended time. A digital simulation of charcoal canister kinetics was also performed to establish limits on how stable radon must be in a space for equilibrium-based calibration to be performed there. Overall, the accuracy of measurements using calibrations from each space differed by less than 10% after 2 d exposure time, showing that carefully controlled conditions are not necessary for the accurate calibration of charcoal canisters. Furthermore, measurement accuracy differed by less than 2% for different spectral region selections. Accuracy improved slightly with canister exposure duration. Simulations suggest that radon instability is most tolerable near the beginning of canister calibration exposures, but this merits further experimental study.

07 ISOTOPE AND RADIATION SOURCES↗

Preliminary Experiences with the Rexon UL-320-FDR: An Automated Thermoluminescent Dosimeter Reader with Removable Contact Heating Planchets and an Infrared Temperature Feedback System

The Rexon UL-320 FDR is a novel resistive-heating thermoluminescent dosimeter reader with a unique temperature measurement system and an automated dosimeter processing mechanism. The removable contact heating planchets have black-body adhesives on the back for capturing temperature information with infrared sensors. A heating cycle feedback loop ensures accurate, precise, and reproducible heating sequences. Heating rates between 0.8 and 40°C s -1 for up to 1,000 s are possible. Photomultiplier tube sensitivity and drift, dark current counts, and planchet glow were measured experimentally. Additionally, 25 LiF:Mg,Ti dosimeters were calibrated to demonstrate reader performance. Sensitivity was optimized at 1,200 V, which produced the highest reference light count to dark current count ratio while extending photomultiplier tube life. Dark current counts measured with typical time-temperature profiles for LiF:Mg,Ti were below 10 counts per channel but increased by up to 2.5% for more extreme heating cycles. Reader sensitivity drifts of up to 10% were observed during extended automated operations with typical time-temperature profiles. Total counts resulting from planchet glow decreased with faster heating rates. Calibrations performed with LiF:Mg,Ti dosimeters yielded results comparable to more established reader designs. Spikes were observed in ~3% of the glow curves from planchet dust and oil burning off at elevated temperatures. The use of N 2 gas and sensitivity drift corrections are recommended to improve dosimetry performance for the UL-320 FDR reader.

61 RADIATION PROTECTION AND DOSIMETRY↗

Simulations and Experimental Verifications of an Algorithm for Radiation Source Mapping and Navigational Path Generation

Accurate and efficient mapping and localization of both ionizing and non-ionizing radiation sources are important across many different fields. As such, a versatile mapping and navigational path generation algorithm, which can be applied to any point source measurements that follow an inverse-square characteristic, was developed using non-linear least squares methods. Forty thousand simulations were performed on the algorithm, which located sources successfully in a 10 m × 10 m × 10 m three-dimensional space with a success rate of over 80% across different noise functions, given a proportional constant of 10 to 1,000. The algorithm was also verified experimentally with small-scale radioactive decontamination of a 70 cm × 70 cm surface and localization of a lost Wi-Fi router in a 70 m × 70 m open field. One hundred twenty-one measurements were taken from each experiment, which were then fed into the algorithm for navigation. For the radioactive 137Cs source, the estimated locations were within 7 cm × 7 cm of the answer in 79.3% of the scenarios, while the Wi-Fi router was located to within 7 m × 7 m in 57.9% of the tests. In general, the method requires much less information and data than a geographically comprehensive survey and thus shows a lot of potential for practical applications, such as lost source retrieval with unmanned aerial vehicles, small-scale decontamination, mapping undocumented Wi-Fi routers or radio towers, and radiation simulation with radio signals. Here, the different failure modes, desirable features, and potential improvements were also identified but remain as future work.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Analysis of Long-Term Quality Control Data for a 137 Cs Dosimetry Calibration Source

Strict quality assurance programs are required for many radiological applications, but these seldom exist for verifying dosimetry calibration sources. After initial characterization of a dosimetry calibration facility, quality control procedures are recommended to ensure the early detection of any changes or malfunctions. These also result in refined knowledge about average dose rate and experimental variations in dose delivery. This paper describes the implementation of a phase I quality control protocol for a 137 Cs dosimetry calibration source and includes an analysis of the resulting data collected over a 24-mo period. During this time, substantial data was collected to establish trial control limits. Air kerma rate measurements were obtained using an ion chamber and were adjusted for decay, corrected for ambient temperature, pressure and humidity, and then analyzed using quality control charts. Three variations of rational subgrouping methods were used in order to find assignable causes of error, and Nelson's Rules were followed to detect any non-random statistical variations. Measurements were subgrouped according to same-day measurements in order to detect positional errors as well as atmospheric correction errors. Additionally, measurements were subgrouped according to analogous experimental setups in order to detect failure in equipment or incorrect settings. Both were analyzed using the X-bar and R chart method. Similarly, individuals and moving ranges charts were used to carefully examine each position in order to observe any situational errors that may occur which include timing, positional, or interference errors. Each method was successful in identifying unique out-of-control data points that occurred during the phase I application of forming control limits. Furthermore, over the 24-mo period, enough data points were deemed in-control to establish reliable trial limits. Future experiments will include the phase II application of gaining more reliable measurements in order to fine-tune the limits, as well as performing a designed experiment, where variables are purposefully changed in order to test the variation of the data.

61 RADIATION PROTECTION AND DOSIMETRY↗

Design and Characterization of an Extremely-Sensitive, Large-Volume Gamma-Ray Spectrometer for Environmental Samples

A large volume gamma spectrometer was designed and constructed to analyze foodstuffs and environmental samples having low radionuclide concentrations. This system uses eight 11-cm × 42.5-cm × 5.5-cm NaI(Tl) detectors, chosen due to their relatively high sensitivity and availability and arranged in an octagonal configuration. The sensitive volume of the system is ~28 cm in diameter and ~42 cm deep. Shielding consists of an 86-cm × 86-cm square, 64-cm-tall lead brick enclosure with 18-cm-thick lead walls lined by 0.3-cm-thick copper plates. An aluminum top was machined to suspend the detectors within this shield. The shielding reduces background counts by 72% at 100 keV and 42% at 1,000 keV. The positional variability in sensitivity of the well was determined by both simulation and experiment. A 2.1-L volume of nearly uniform sensitivity, varying less than 10%, exists in the well's center. Energy resolutions of 14.6% and 7.8% were measured for 241 Am and 137 Cs, respectively. Energy resolution shows a 0.2% variation for both 241 Am and 137 Cs as a function of position within all regions of the well’s central sensitive volume. Dead time was also determined to be less than 35% for all sources measured in the system, the largest of which had an activity of 1,760 kBq. Simulated results for various source geometries show higher counts for smaller samples, especially at lower energies due to less attenuation of low energy photons. In conclusion, minimum detectable activities were determined for all source energies used, less than 5.1 Bq kg -1 for reasonable background and sample counting times.

47 OTHER INSTRUMENTATION↗