Development of an L-Edge X-ray Absorbance Spectrometer for Monitoring Dissolver Solutions in H-Canyon

Savannah River National Laboratory has developed a monitor to measure plutonium and uranium concentrations in solutions of dissolved nuclear fuel. The monitor will be installed in the sample aisle location for the 6.3D Dissolver in the Savannah River Site’s H-Canyon and used in support of the electrolytic dissolution such as Fast Critical Assembly fuel. The monitor is based on the atomic absorbance of x-rays. Elements are differentiated by the appearance of absorbance features at specific energies of the x-ray spectrum that correspond to L-edge transitions of inner core electrons. Hence, the technique is called L-Edge X-Ray Absorbance Spectroscopy (L-XRAS). The technique is suitable for nuclear fuel processing due to its relative insensitivity to other components of the dissolver solution, such as nitric acid, transition metals (Fe, Cr, Ni, Mn) such as those from stainless steel, particulates, and catalysts and additives. The instrumentation consists of a commercially available x-ray source and detector, a sample cell designed to interface with the airlift sampler associated with H-Canyon Tank 6.3D, and a stainless steel enclosure. SRNL wrote instrument control software and developed chemometric models to interpret x-ray intensity spectra and estimate analyte concentrations and uncertainties in real time.

12 MANAGEMENT OF RADIOACTIVE AND NON-RADIOACTIVE W↗

Instrument Control Software for Spectrophotometers

SRNL has developed instrument control software for diode array spectrophotometers that can be used for real time monitoring of dynamic systems. This software has evolved through several iterations as instruments have been developed for use in processing facilities at Savannah River Site. There are several noteworthy features of the software. The wavelength calibration is checked and updated in real time against reference emission lines that are designed into the instrument light source. Nonlinearities and offsets associated with the intensity response of the instruments are characterized, and absorbance spectra are appropriately corrected. These actions are necessary for accurate calculation of an absorbance spectrum from the intensity responses of two individual diode array detectors. The corrections also result in the spectra obtained for a particular sample by two (or more) different diode array detector pairs being nearly identical. This property assures consistent measurements at multiple locations with different instruments. It also permits the use of a single analyte calibration across many instruments without modification and without loss of accuracy

47 OTHER INSTRUMENTATION↗

H-Canyon Dissolver Offgas Monitoring Using Raman Spectroscopy

This report describes the development and in facility demonstration of a Raman spectrometer to monitor known aluminum offgas components generated during the dissolution of Spent Nuclear Fuel at the Savannah River Site H-Canyon facility. The presence of these gases is an indicator that material is actively being dissolved. This process monitoring technique, which measures the offgas in real time, has the potential to improve the efficiency of facility operations. If the dissolution time is too short, time is wasted due to extended dissolution periods with additional probing of the dissolver pot to confirm that fragment levels are below the thresholds for continuing to the next charge or moving the dissolvent to a receiving tank. If the dissolution time is too long, time and energy are wasted.

12 MANAGEMENT OF RADIOACTIVE AND NON-RADIOACTIVE W↗

Dissolution Flowsheet for U-10Mo Scrap Generated from the Fabrication of High Performance Research Reactor Fuel

The Savannah River National Laboratory was requested to develop dissolution flowsheets for low enriched uranium scrap streams generated during the fabrication of high performance research reactor (HPRR)fuel. The scrap streams include U-10Mo-Zr foil and rejected Al-clad plates. Co-rolled U-10Mo-Zr foils and Alclad U-10Mo-Zr mini-plates which have been through the hot isostatic pressing process were obtained from BWX Technologies, Inc. for use in this work. Small-scale dissolution experiments were performed to identify the flowsheet conditions required to safely and efficiently dissolve the fuel fabrication scrap for subsequent U recovery. Experiments were performed to demonstrate the dissolution of the U-10Mo-Zr foil, removal of the Al cladding from the mini-plates, and the dissolution of the mini-plates following cladding removal.

11 NUCLEAR FUEL CYCLE AND FUEL MATERIALS↗

Spectrophotometers for Uranium Monitoring in H Canyon

SRNL Analytical Development and R&D Engineering have developed an On-Line Monitoring Spectrophotometer to provide measurements of uranium and nitric acid at several locations in H Canyon Second Uranium Cycle. Instruments are installed at Tanks 15.4, 16.8, 17.4, 17.5, and 18.7. Two-sigma uncertainties for the instruments are 5.3% uranium (all tanks) and 11% nitric acid (Tanks 16.8 and 17.5). The spectrophotometers are of similar design to those recently installed in HB-Line for Pu monitoring, although some changes have been made to make them more suitable for uranium measurements. A full description of design, components, and operating procedures is given.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Dissolution of Material Test Reactor Fuel in an H-Canyon Dissolver

In an amended record of decision for the management of spent nuclear fuel (SNF) at the Savannah River Site, the US Department of Energy has authorized the dissolution and recovery of U from 1000 bundles of Al-clad SNF. The SNF is fuel from domestic and foreign research reactors and is typically referred to as Material Test Reactor (MTR) fuel. Bundles of MTR fuel containing assemblies fabricated from U-Al alloys (or other U compounds) are currently dissolved using a Hg-catalyzed HNO 3 flowsheet. Since the development of the existing flow sheet, improved experimental methods have been developed to more accurately characterize the offgas composition and generation rate during laboratory dissolutions. Recently, these new techniques were successfully used to develop a flowsheet for the dissolution of High Flux Isotope Reactor (HFIR) fuel. Using the data from the HFIR dissolution flowsheet development and necessary laboratory experiments, the Savannah River National Laboratory (SRNL) was requested to define flowsheet conditions for the dissolution of MTR fuels. With improved offgas characterization techniques, SRNL will be able define the number of bundles of fuel which can be charged to an H-Canyon dissolver with much less conservatism.

12 MANAGEMENT OF RADIOACTIVE AND NON-RADIOACTIVE W↗

Dissolution of Material and Test reactor Fuel in an H-Canyon Dissolver

In an amended record of decision for the management of spent nuclear fuel (SNF) at the Savannah River Site, the US Department of Energy has authorized the dissolution and recovery of U from 1000 bundles of Al-clad SNF. The SNF is fuel from domestic and foreign research reactors and is typically referred to as Material Test Reactor (MTR) fuel. Bundles of MTR fuel containing assemblies fabricated from U-Al alloys (or other U compounds) are currently dissolved using a Hg-catalyzed HNO3 flowsheet. Since the development of the existing flowsheet, improved experimental methods have been developed to more accurately characterize the offgas composition and generation rate during laboratory dissolutions. Recently, these new techniques were successfully used to develop a flowsheet for the dissolution of High Flux Isotope Reactor (HFIR) fuel. Using the data from the HFIR dissolution flowsheet development and necessary laboratory experiments, the Savannah River National Laboratory (SRNL) was requested to define flowsheet conditions for the dissolution of MTR fuels. With improved offgas characterization techniques, SRNL will be able define the number of bundles of fuel which can be charged to an H-Canyon dissolver with much less conservatism.

11 NUCLEAR FUEL CYCLE AND FUEL MATERIALS↗

Dissolution Flowsheet for High Flux Isotope Reactor Fuel

As part of the Spent Nuclear Fuel (SNF) processing campaign, H-Canyon is planning to begin dissolving High Flux Isotope Reactor (HFIR) fuel in late FY17 or early FY18. Each HFIR fuel core contains inner and outer fuel elements which were fabricated from uranium oxide (U 3 O 8 ) dispersed in a continuous Al phase using traditional powder metallurgy techniques. Fuels fabricated in this manner, like other SNF’s processed in H-Canyon, dissolve by the same general mechanisms with similar gas generation rates and the production of H 2 . The HFIR fuel cores will be dissolved and the recovered U will be down-blended into low-enriched U. HFIR fuel was previously processed in H-Canyon using a unique insert in both the 6.1D and 6.4D dissolvers. Multiple cores will be charged to the same dissolver solution maximizing the concentration of dissolved Al. The objective of this study was to identify flowsheet conditions through literature review and laboratory experimentation to safely and efficiently dissolve the HFIR fuel in H-Canyon. Laboratory-scale experiments were performed to evaluate the dissolution of HFIR fuel using both Al 1100 and Al 6061 T6 alloy coupons. The Al 1100 alloy was considered a representative surrogate which provided an upper bound on the generation of flammable (i.e., H 2 ) gas during the dissolution process. The dissolution of the Al 6061 T6 alloy proceeded at a slower rate than the Al 1100 alloy, and was used to verify that the target Al concentration in solution could be achieved for the selected Hg concentration. Mass spectrometry and Raman spectroscopy were used to provide continuous monitoring of the concentration of H 2 and other permanent gases in the dissolution offgas, allowing the development of H 2 generation rate profiles. The H 2 generation rates were subsequently used to evaluate if a full HFIR core could be dissolved in an H-Canyon dissolver without exceeding 60% of the calculated lower flammability limit (LFL) for H 2 at a given Hg concentration. Complete dissolution of the Al 1100 and Al 6061 T6 alloys up to a final Al concentration of 2 M was obtained using a 7 M HNO 3 solution containing a 0.002 M Hg catalyst. However, following the dissolutions, solids were observed in the solution. The solids were amorphous, but likely originated from the Si present in the alloys. No crystalline materials, such as Al(NO 3 ) 3 were observed. During the course of the dissolution experiments, it was determined that delaying the addition of Hg once the HNO 3 solution reached the boiling point can reduce the total offgas and H 2 generation rates. The delay in starting the Hg addition is not necessary for HFIR fuel dissolution, but could be useful in other research reactor dissolution campaigns. The potential to generate flammable concentrations of H 2 in the offgas during a HFIR fuel dissolution was evaluated using the experimental data. The predicted H 2 concentration in the dissolver offgas stream was compared with 60% of the calculated H 2 LFL at 200 °C using several prototypical experiments. The calculations showed that a full HFIR core can be dissolved using nominally 0.002 M Hg to catalyze the dissolution. The margin between the predicted H 2 concentration and the calculated LFL was greater when the solution was allowed to boil for 45 min prior to initiating the Hg addition. When the Hg was increased to 0.004 M, the predicted H 2 concentration exceeded the calculated LFL early in the dissolution. The dissolution experiments also demonstrated that additional Hg (beyond the initial 0.002 M) could be added as the Al concentration increases. The ability to add more Hg during a HFIR fuel dissolution could be beneficial if slow dissolution rates are observed at high Al concentrations. Experimental data were used to demonstrate that the predicted H 2 concentration in a dissolver was below 60% of the calculated LFL at 200 °C when 0.004 M Hg was used to catalyze the dissolution if the Al concentration is conservatively greater than 0.5 M. Data also show that the Hg concentration during a HFIR fuel dissolution can be increased from 0.002 to 0.008 M at an Al concentration of 1.3 M.

11 NUCLEAR FUEL CYCLE AND FUEL MATERIALS↗

Spectrophotometers for plutonium monitoring in HB-line

This report describes the equipment, control software, calibrations for total plutonium and plutonium oxidation state, and qualification studies for the instrument. It also provides a detailed description of the uncertainty analysis, which includes source terms associated with plutonium calibration standards, instrument drift, and inter-instrument variability. Also included are work instructions for instrument, flow cell, and optical fiber setup, work instructions for routine maintenance, and drawings and schematic diagrams.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Summary of recent Raman Spectroscopy testing of SRS processes

This report describes several scoping projects conducted at SRNL using Raman spectroscopic methods for monitoring different aspects of nuclear waste and materials processing. One project examined the suitability of a Raman telescope for in situ measurement of solid residues in waste tanks. Characteristics evaluated for this equipment included radiation resistance, ease of use, and sensitivity. A second project monitored the nitrate content in liquid filtrate from the testing of a rotary microfilter using a fiber-based insertion probe. The third project made Raman measurements of various gases, including H 2 and NO x , in the headspace of a vessel while dissolving aluminum coupons in nitric acid. Measurements followed the evolution of these species in real time. Although the majority of these projects occurred in the laboratory environment, SRNL has substantial experience with implementing other optical techniques into nuclear materials processing environments. The work described in this report shows the potential of the Raman technology to provide real time measurements of species such as nitrate or hydroxide during sludge washing or evolved gases such as hydrogen or NO x during waste processing.

12 MANAGEMENT OF RADIOACTIVE AND NON-RADIOACTIVE W↗

Interpretation of At-Line Spectra from AFS-2 Batch #3 Ferrous Sulfamate Treatment

Spectra from the “at-line” spectrometer were obtained during the ferrous sulfamate (FS) valence adjustment step of AFS-2 Batch #3 on 9/18/2013. These spectra were analyzed by mathematical principal component regression (PCR) techniques to evaluate the effectiveness of this treatment. Despite the complications from Pu(IV), we conclude that all Pu(VI) was consumed during the FS treatment, and that by the end of the treatment, about 85% was as Pu(IV) and about 15% was as Pu(III). Due to the concerns about the “odd” shape of the Pu(IV) peak and the possibility of this behavior being observed in the future, a follow-up sample was sent to SRNL to investigate this further. Analysis of this sample confirmed the previous results and concluded that it “odd” shape was due to an intermediate acid concentration. Since the spectral evidence shows complete reduction of Pu(VI) we conclude that it is appropriate to proceed with processing of this the batch of feed solution for HB-Line including the complexation of the fluoride with aluminum nitrate.

38 RADIATION CHEMISTRY, RADIOCHEMISTRY, AND NUCLEA↗

Plutonium loading capacity of reillex™ hpq anion exchange column - AFS-2 plutonium flowsheet for mox

Radioactive plutonium (Pu) anion exchange column experiments using scaled HB-Line designs were performed to investigate the dependence of column loading performance on the feed composition in the H-Canyon dissolution process for plutonium oxide (PuO 2 ) product shipped to the Mixed Oxide (MOX) Fuel Fabrication Facility (MFFF). These loading experiments show that a representative feed solution containing ~5 g Pu/L can be loaded onto Reillex™ HPQ resin from solutions containing 8 M total nitrate and 0.1 M KF provided that the F is complexed with Al to an [Al]/[F] molar ratio range of 1.5-2.0. Lower concentrations of total nitrate and [Al]/[F] molar ratios may still have acceptable performance but were not tested in this study. Loading and washing Pu losses should be relatively low (<1%) for resin loading of up to 60 g Pu/L. Loading above 60 g Pu/L resin is possible, but Pu wash losses will increase such that 10-20% of the additional Pu fed may not be retained by the resin as the resin loading approaches 80 g Pu/L resin.

11 NUCLEAR FUEL CYCLE AND FUEL MATERIALS↗

Scanning Electron Microscopy and X-Ray Diffraction Analysis of Tank 18 Samples

The F-Area Tank Farm (FTF) Performance Assessment (PA) utilizes waste speciation in the waste release model used in the FTF fate and transport modeling. The waste release modeling associated with the residual plutonium in Tank 18 has been identified as a primary contributor to the Tank 18 dose uncertainty. In order to reduce the uncertainty related to plutonium in Tank 18, a better understanding of the plutonium speciation in the Tank 18 waste (including the oxidation state and stoichiometry) is desired. Savannah River National Laboratory (SRNL) utilized Scanning Electron Microscopy (SEM) and X-ray Diffraction (XRD) to analyze Tank 18 samples to provide information on the speciation of plutonium in the waste material. XRD analysis of the Tank 18 samples did not identify any plutonium mineral phases in the samples. These indicates the crystalline mineral phases of plutonium are below the detection limits of the XRD method or that the plutonium phase(s) lack long range order and are present as amorphous or microcrystalline solids. SEM analysis of the Tank 18 samples did locate particles containing plutonium. The plutonium was found as small particles, usually <1 μm but ranging up to several micrometers in diameter, associated with particles of an iron matrix and at low concentration in other elemental matrices. This suggests the plutonium has an affinity for the iron matrix. Qualitatively, the particles of plutonium found in the SEM analysis do not appear to account for all of the plutonium in the sample based on concentrations determined from the chemical analysis of the Tank 18 samples. This suggests that plutonium is also distributed throughout the solids in low concentrations.