Search NASA⌕ Search

Engineering topics

Overman, Cory T.

Publications and source records attributed to Overman, Cory T..

27 records · Page 2

Controlling Low-Rate Signal Path Microdischarge for an Ultra-Low-Background Proportional Counter

ABSTRACT Pacific Northwest National Laboratory (PNNL) has developed an ultra-low-background proportional counter (ULBPC) made of high purity copper. These detectors are part of an ultra-low-background counting system (ULBCS) in the newly constructed shallow underground laboratory at PNNL (at a depth of ~30 meters water-equivalent). To control backgrounds, the current preamplifier electronics are located outside the ULBCS shielding. Thus the signal from the detector travels through ~1 meter of cable and is potentially susceptible to high voltage microdischarge and other sources of electronic noise. Based on initial successful tests, commercial cables and connectors were used for this critical signal path. Subsequent testing across different batches of commercial cables and connectors, however, showed unwanted (but still low) rates of microdischarge noise. To control this noise source, two approaches were pursued: first, to carefully validate cables, connectors, and other commercial components in this critical signal path, making modifications where necessary; second, to develop a custom low-noise, low-background preamplifier that can be integrated with the ULBPC and thus remove most commercial components from the critical signal path. This integrated preamplifier approach is based on the Amptek A250 low-noise charge-integrating preamplifier module. The initial microdischarge signals observed are presented and characterized according to the suspected source. Each of the approaches for mitigation is described, and the results from both are compared with each other and with the original performance seen with commercial cables and connectors.

low noise↗

Digital Pulse-Shape Discrimination Applied to an Ultra-Low-Background Gas-Proportional Counting System: First Results

Abstract A new ultra-low-background proportional counter (ULBPC) design was recently developed at Pacific Northwest National Laboratory (PNNL). This design, along with an ultra-low-background counting system (ULBCS) which provides passive and active shielding with radon exclusion, has been developed to complement a new shallow underground laboratory (~30 meters water-equivalent) constructed at PNNL. After these steps to mitigate dominant backgrounds (cosmic rays, external gamma-rays, radioactivity in materials), remaining background events do not exclusively arise from ionization of the proportional counter gas. Digital pulse-shape discrimination (PSD) is thus employed to further improve measurement sensitivity. In this work, a template shape is generated for each individual sample measurement of interest, a "self-calibrating" template. Differences in event topology can also cause differences in pulse shape. In this work, the temporal region analyzed for each event is refined to maximize background discrimination while avoiding unwanted sensitivity to event topology. This digital PSD method is applied to sample and background data, and initial measurement results from a biofuel methane sample are presented in the context of low-background measurements currently being developed.

signal processing, pulse-shape discrimination, pro↗

The Design, Construction, and Initial Characterization of an Ultra-Low-Background Gas-Proportional Counting System

ABSTRACT Over the past several years, the Pacific Northwest National Laboratory (PNNL) has developed an ultra-low background proportional counter (ULBPC) technology. The resulting detector is the product of an effort to produce a low-background, physically robust gas proportional counter for applications like radon emanation measurements, groundwater tritium, and 37Ar. In order to fully take advantage of the inherent low-background properties designed into the ULBPC, a comparably low-background dedicated counting system is required. An ultra-low-background counting system (ULBCS) was recently built in the new shallow underground laboratory at PNNL. With a design depth of 30 meters water-equivalent, the shallow underground laboratory provides approximately 100x fewer fast neutrons and 6x fewer muons than a surface location. The ULBCS itself provides additional shielding in the form of active anti-cosmic veto (via 2-in. thick plastic scintillator paddles) and passive borated poly (1 in.), lead (6 in.), and copper (~3 in.) shielding. This work will provide details on PNNL’s new shallow underground laboratory, examine the motivation for the design of the counting system, and provide results from the characterization of the ULBCS, including initial detector background.

ultra-low-background, proportional counters, shall↗

Empirical Correction of Crosstalk in a Low-Background Germanium γ–γ Analysis System

ABSTRACT The Pacific Northwest National Laboratory is currently developing a custom software suite capable of automating many of the tasks required to accurately analyze coincident signals within gamma spectrometer arrays. During the course of this work, significant crosstalk was identified in the energy determination for spectra collected with a new low-background intrinsic germanium (HPGe) array at PNNL. The HPGe array is designed for high detection efficiency, ultra-low-background performance, and sensitive gamma gamma coincidence detection. The first half of the array, a single cryostat containing 7 HPGe crystals, was recently installed into a new shallow underground laboratory facility. This update will present a brief review of the germanium array, describe the observed crosstalk, and present a straight-forward empirical correction that significantly reduces the impact of this crosstalk on the spectroscopic performance of the system.

coincidence↗

Construction of a Shallow Underground Low-background Detector for a CTBT Radionuclide Laboratory

The International Monitoring System (IMS) is a verification component of the Comprehensive Nuclear-Test-Ban Treaty (CTBT), and in addition to a series of radionuclide monitoring stations, contains sixteen radionuclide laboratories capable of verification of radionuclide station measurements. This paper presents an overview of a new commercially obtained low-background detector system for radionuclide aerosol measurements recently installed in a shallow (>30 meters water equivalent) underground clean-room facility at Pacific Northwest National Laboratory. Specifics such as low-background shielding materials, active shielding methods, and improvements in sensitivity to IMS isotopes will be covered.

Forrester, Joel B.↗

Cryostat for Ultra-low-energy Threshold Germanium Spectrometers

This paper presents progress on the development of a cryostat intended to improve upon the low-energy threshold (below 0.5 keV) of p-type point contact germanium gamma-ray spectrometers. Ultra-low energy thresholds are important in the detection of low-energy nuclear recoils, an event class relevant to both dark matter direct detection and measurement of coherent neutrino-nucleus scattering. The cryostat design, including a thermal and electrical-field model, is presented. A prototype cryostat has been assembled and data acquired to evaluate its vacuum and thermal performance.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

The C-4 Dark Matter Experiment

Abstract We describe the experimental design of C-4, an expansion of the CoGeNT dark matter search to four identical detectors each approximately three times the mass of the p-type point contact (PPC) germanium diode presently taking data at the Soudan Underground Laboratory. Expected reductions of radioactive backgrounds and energy threshold are discussed, including an estimate of the additional sensitivity to low-mass dark matter candidates to be obtained with this search.

72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS↗

A Germanium Array for Sample Measurements: Recent Results and New Analysis Methods

ABSTRACT The Pacific Northwest National Laboratory is finalizing construction of a 14-crystal intrinsic germanium (HPGe) array designed for high detection efficiency, ultra-low-background performance, and sensitive gamma gamma coincidence detection. Initial development of a custom software suite capable of automating many of the tasks required to accurately analyze coincident signals within close-geometry detector arrays will also be completed this year. The analysis capability is required to enable rapid interpretation of data collected with the low-background (HPGe) array. The status of final construction of the 2nd cryostat and the complete system will be discussed, as well as measurement results from independent operation of the 1st cryostat. Status of the CASCADES analysis development, including cross-talk correction and new techniques for efficiency calibration, are also presented.

Keillor, Martin E.↗

Majorana Electroformed Copper Mechanical Analysis

The MAJORANA DEMONSTRATOR is a large array of ultra-low background high-purity germanium detectors, enriched in 76 Ge, designed to search for zero-neutrino double-beta decay. The DEMONSTRATOR will utilize ultra high purity electroformed copper for a variety of detector components and shielding. A preliminary mechanical evaluation was performed on the Majorana prototype electroformed copper material. Several samples were removed from a variety of positions on the mandrel. Tensile testing, optical metallography, scanning electron microscopy, and hardness testing were conducted to evaluate mechanical response. Analyses carried out on the Majorana prototype copper to this point show consistent mechanical response from a variety of test locations. Evaluation shows the copper meets or exceeds the design specifications.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗