First Tests for the Detection of the LINAC Irradiation Field Using PIN Diodes
The employment of the technology of semiconductor detectors, in the medical physics environment is of great importance due to its versatility and dependability. In this work we present the first results and the experimental arrangement employed with PIN diodes that are conditioned for the measurement of the field of irradiation of a lineal accelerator (LINAC) used in radiotherapy. In our tests we used a PIN photodiode. In former experiments, this diode presented a response to the intensity of the applied field when it was exposed to an X-ray beam in medical and industrial radiography equipments. This diode is a low cost and easy acquisition one in the field. These characteristics transform it into a serious candidate as detector to be used in electronic arrangements for the detection of radiation fields in radio-therapy with X-rays. Experiments were designed to obtain the response of this diode when it was exposed to X-ray beams of a LINAC used in radiotherapy. Firstly the tests were carried out for a 6 MeV photon beam with a source to surface distance (SSD) of 100 cm, obtaining very encouraging results. We seek to carry out tests for more energy values in order to obtain the energy response of this detector as a radiation sensor device. This device could be applied in the design of working tools, for example, for the quality control in procedures of radiotherapy.