Filter Test System for Nuclear Container Filters
The purpose of this procedure is to measure the collection efficiency of the filters that are integrated into the lids of containers for nuclear material at Los Alamos National Laboratory (LANL). As an application of this procedure, a filter test report certificate can be created to document the measurement process. This procedure is intended to describe the TA-55, PF-4 (room 6A) operation of a Filter Test System (FTS) for Hagan and SAVY storage containers.