Search NASA⌕ Search

Engineering topics

Rehak, P

Publications and source records attributed to Rehak, P.

ASIC for SDD-Based X-Ray Spectrometers

We present an application-specific integrated circuit (ASIC) for high-resolution x-ray spectrometers (XRS). The ASIC reads out signals from pixelated silicon drift detectors (SDDs). The pixel does not have an integrated field effect transistor (FET); rather, readout is accomplished by wire-bonding the anodes to the inputs of the ASIC. The ASIC dissipates 32 mW, and offers 16 channels of low-noise charge amplification, high-order shaping with baseline stabilization, discrimination, a novel pile-up rejector, and peak detection with an analog memory. The readout is sparse and based on custom low-power tristatable low-voltage differential signaling (LPT-LVDS). A unit of 64 SDD pixels, read out by four ASICs, covers an area of 12.8 cm{sup 2} and dissipates with the sensor biased about 15 mW/cm{sup 2}. As a tile-based system, the 64-pixel units cover a large detection area. Our preliminary measurements at -44 C show a FWHM of 145 eV at the 5.9 keV peak of a {sup 55}Fe source, and less than 80 eV on a test-pulse line at 200 eV.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Charge-Pump Detector for X-ray Correlation Spectroscopy

A detector for the X-ray Correlation Spectroscopy (XCS) instrument at the Linac Coherent Light Source (LCLS) in Stanford (CA) is being developed at Brookhaven National Laboratory (BNL). The LCLS is the first operational X-ray free electron laser. It provides extremely bright coherent laser-like X-ray pulses with energy up to 8 keV, shorter than 100 fs and with a repetition rate that will go up to 120 Hz. An ideal detector for XCS experiments should cover a large angular range with high efficiency and provide a proper resolution to resolve the speckle. The requirement for dynamic range is not particularly stringent while a fast readout is needed. In particular, the Charge Pump Detector has to be highly efficient at the energy of 8 keV, provide a dynamic range of 100 photons and a readout noise much better than one photon. The 1024 x 1024 pixels have to be read within the repetition rate of the laser pulses, that is faster than 8 ms. The pixel size of 56 {micro} m x 56 {micro}m is a compromise between charge sharing and small pixel. Working principle and details of the detector will be discussed.

43 PARTICLE ACCELERATORS↗

R&D Towards Cryogenic Optical Links

A number of critical active and passive components of optical links have been tested at 77 K or lower temperatures, demonstrating potential development of optical links operating inside the liquid argon time projection chamber (LArTPC) detector cryostat. A ring oscillator, individual MOSFETs, and a high speed 16:1 serializer fabricated in a commercial 0.25-{micro}m silicon-on-sapphire CMOS technology continued to function from room temperature to 4.2 K, 15 K, and 77 K respectively. Three types of laser diodes lase from room temperature to 77 K. Optical fibers and optical connectors exhibited minute attenuation changes from room temperature to 77 K.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

ASIC for SDD-Based X-ray Spectrometers

We present an application-specific integrated circuit (ASIC) for high-resolution x-ray spectrometers (XRS). The ASIC reads out signals from pixelated silicon drift detectors (SDDs). The pixel does not have an integrated field effect transistor (FET); rather, readout is accomplished by wire-bonding the anodes to the inputs of the ASIC. The ASIC dissipates 32 mW, and offers 16 channels of low-noise charge amplification, high-order shaping with baseline stabilization, discrimination, a novel pile-up rejector, and peak detection with an analog memory. The readout is sparse and based on custom low-power tristatable low-voltage differential signaling (LPT-LVDS). A unit of 64 SDD pixels, read out by four ASICs, covers an area of 12.8 cm{sup 2} and dissipates with the sensor biased about 15 mW/cm{sup 2}. As a tile-based system, the 64-pixel units cover a large detection area. Our preliminary measurements at -44 C show a FWHM of 145 eV at the 5.9 keV peak of a {sup 55}Fe source, and less than 80 eV on a test-pulse line at 200 eV.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Development of Thin-Window Silicon Drift Detector for X-ray Spectroscopy

A new set of thin-window silicon drift detectors composed of an array of hexagonal shaped detectors has been designed, constructed and tested for X-ray spectroscopy. Each individual ThinWinSDD has a thin entrance window on one side and a spiral shaped hexagonal cathode around a center anode on the other side. To produce the thin entrance window a 10 keV implantation of boron through a 500 A silicon dioxide was used. The implantation was followed by an annealing at 700 C for 30 min and a reactive ion etching step to ensure the removal of silicon dioxide from the smallest feature (5 mum). An aluminum layer is coated in the same vacuum system after back-sputtering. This step involves removing the native oxide that has formed on the top of the silicon substrate and then sputtering a 1100 A thick layer of aluminum onto the X-ray entrance window. The aluminum layer must be thick enough to block visible light, but thin enough to be transparent to soft X-rays down to 280 eV. We discuss first test results that include detector leakage current measurements and the response for multiple detectors exposed to the National Synchrotron Light Source's UV beam line U3C located at Brookhaven National Laboratory for X-ray energies as low as 280 eV.

36 MATERIALS SCIENCE↗

Development of a Silicon Drift Detector Array: An X-ray Fluorescence Spectrometer for Remote Surface Mapping

Over the past three years NASA Marshall Space Flight Center has been collaborating with Brookhaven National Laboratory to develop a modular Silicon Drift Detector (SDD) X-Ray Spectrometer (XRS) intended for fine surface mapping of the light elements of the moon. The value of fluorescence spectrometry for surface element mapping is underlined by the fact that the technique has recently been employed by three lunar orbiter missions; Kaguya, Chandrayaan-1, and Chang'e. The SDD-XRS instrument we have been developing can operate at a low energy threshold (i.e. is capable of detecting Carbon), comparable energy resolution to Kaguya (<150 eV at 5.9 keV) and an order of magnitude lower power requirement, making much higher sensitivities possible. Furthermore, the intrinsic radiation resistance of the SDD makes it useful even in radiation-harsh environments such as that of Jupiter and its surrounding moons.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

XAMPS Detectors Readout ASIC for LCLS

An ASIC for the readout of signals from X-ray Active Matrix Pixel Sensor (XAMPS) detectors to be used at the Linac Coherent Light Source (LCLS) is presented. The X-ray Pump Probe (XPP) instrument, for which the ASIC has been designed, requires a large input dynamic range on the order of 104 photons at 8 keV with a resolution of half a photon FWHM. Due to the size of the pixel and the length of the readout line, large input capacitance is expected, leading to stringent requirement on the noise optimization. Furthermore, the large number of pixels needed for a good position resolution and the fixed LCLS beam period impose limitations on the time available for the single pixel readout. Considering the periodic nature of the LCLS beam, the ASIC developed for this application is a time-variant system providing low-noise charge integration, filtering and correlated double sampling. In order to cope with the large input dynamic range a charge pump scheme implementing a zero-balance measurement method has been introduced. It provides an on chip 3-bit coarse digital conversion of the integrated charge. The residual charge is sampled using correlated double sampling into analog memory and measured with the required resolution. The first 64 channel prototype of the ASIC has been fabricated in TSMC CMOS 0.25 {micro}m technology. In this paper, the ASIC architecture and performances are presented.

43 PARTICLE ACCELERATORS↗

Performance of Thin-Window Silicon Drift Detectors

Several sets of hexagonal Silicon Drift Detector (SDD) arrays were produced at BNL and by a commercial vendor, KETEK. Each array consists of 14 independent detectors (pixels) and two additional test pixels at two of the corners. The side of the detector upon which the X-ray radiation is incident (window side) has a thin junction covering the entire active area. The opposite side (device side) contains a drift-field electrode structure in the form of a hexagonal spiral and an electron collecting anode. There are 4 guard rings surrounding the 14-pixel array area on both sides of the detector. Within each array, 7 of the pixels have an aluminum field plate - interrupted spirals that stabilize the electric potential under the Si-SiO2 interface, while the other 7 do not. The drift field in the silicon volume is controlled by three biases: one is applied to a rectifying contact, one to the detector entrance window, and the third to a contact on the outer portion of the spiral common to all pixels in the array. Some arrays have been newly measured in NSLS beam line U3C at BNL. The complete assemblies were installed in the vacuum and cooled to ?27 C. During this run, spectra for energies ranging between 400 and 900 eV were collected in several pixels, some with field plates and others without. The detailed testing results of several arrays are reported here.

72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS↗

Development of Thin-Window Silicon Drift Detector for X-ray Spectroscopy

A new set of Thin-Window Silicon Drift Detectors composed of an array of hexagonal shaped detectors has been designed, constructed and tested for X-ray spectroscopy. Each individual ThinWinSDD has a thin entrance window on one side and a spiral shaped hexagonal cathode around a center anode on the other side. To produce the thin entrance window a 10 keV implantation of boron through a 500 {angstrom} silicon dioxide was used. The implantation was followed by an annealing at 700 C for 30 min and a reactive ion etching step to ensure the removal of silicon dioxide from the smallest feature (5 {micro}m). An aluminum layer is coated in the same vacuum system after back-sputtering. This step involves removing the native oxide that has formed on the top of the silicon substrate and then sputtering a 1100 {angstrom} thick layer of aluminum onto the X-ray entrance window. The aluminum layer must be thick enough to block visible light, but thin enough to be transparent to soft x-rays down to 280 eV. We discuss first test results that include detector leakage current measurements and the response for multiple detectors exposed to the National Synchrotron Light Source's UV beam line U3C located at Brookhaven National Laboratory for X-ray energies as low as 280 eV.

36 MATERIALS SCIENCE↗

Differential Phase Contrast with a Segmented Detector in a Scanning X-ray Microprobe

Scanning X-ray microprobes are unique tools for the nanoscale investigation of specimens from the life, environmental, materials and other fields of sciences. Typically they utilize absorption and fluorescence as contrast mechanisms. Phase contrast is a complementary technique that can provide strong contrast with reduced radiation dose for weakly absorbing structures in the multi-keV range. In this paper the development of a segmented charge-integrating silicon detector which provides simultaneous absorption and differential phase contrast is reported. The detector can be used together with a fluorescence detector for the simultaneous acquisition of transmission and fluorescence data. It can be used over a wide range of photon energies, photon rates and exposure times at third-generation synchrotron radiation sources, and is currently operating at two beamlines at the Advanced Photon Source. Images obtained at around 2 keV and 10 keV demonstrate the superiority of phase contrast over absorption for specimens composed of light elements.

43 PARTICLE ACCELERATORS↗