Multi-energy reconstructions, central electron temperature measurements, and early detection of the birth and growth of runaway electrons using a versatile soft x-ray pinhole camera at MST

A multi-energy soft x-ray pinhole camera has been designed, built, and deployed at the Madison Symmetric Torus to aid the study of particle and thermal transport, as well as MHD stability physics. This novel imaging diagnostic technique employs a pixelated x-ray detector in which the lower energy threshold for photon detection can be adjusted independently on each pixel. The detector of choice is a PILATUS3 100 K with a 450 μm thick silicon sensor and nearly 100 000 pixels sensitive to photon energies between 1.6 and 30 keV. An ensemble of cubic spline smoothing functions has been applied to the line-integrated data for each time-frame and energy-range, obtaining a reduced standard-deviation when compared to that dominated by photon-noise. The multi-energy local emissivity profiles are obtained from a 1D matrix-based Abel-inversion procedure. Central values of Te can be obtained by modeling the slope of the continuum radiation from ratios of the inverted radial emissivity profiles over multiple energy ranges with no a priori assumptions of plasma profiles, magnetic field reconstruction constraints, high-density limitations, or need of shot-to-shot reproducibility. In tokamak plasmas, a novel application has recently been tested for early detection, 1D imaging, and study of the birth, exponential growth, and saturation of runaway electrons at energies comparable to 100 × Te,0; thus, early results are also presented.

Delgado-Aparicio, L. F. (ORCID:0000000167394380)↗

Microcalorimeter measurement of x-ray spectra from a high-temperature magnetically confined plasma

A NASA-built x-ray microcalorimeter spectrometer has been installed on the MST facility at the Wisconsin Plasma Physics Laboratory and has recorded x-ray photons emitted by impurity ions of aluminum in a majority deuterium plasma. Much of the x-ray microcalorimeter development has been driven by the needs of astrophysics missions, where imaging arrays with few-eV spectral resolution are required. The goal of our project is to adapt these single-photon-counting microcalorimeters for magnetic fusion energy research and demonstrate the value of such measurements for fusion science. Microcalorimeter spectrometers combine the best characteristics of the x-ray instrumentation currently available on fusion devices: high spectral resolution similar to an x-ray crystal spectrometer and the broadband coverage of an x-ray pulse height analysis system. Fusion experiments are increasingly employing high-Z plasma-facing components and require measurement of the concentration of all impurity ion species in the plasma. This diagnostic has the capability to satisfy this need for multi-species impurity ion data and will also contribute to measurements of impurity ion temperature and flow velocity, Zeff, and electron density. Here, we introduce x-ray microcalorimeter detectors and discuss the diagnostic capability for magnetic fusion energy experiments. We describe our experimental setup and spectrometer operation approach at MST, and we present the results from an initial measurement campaign.

Eckart, M. E. (ORCID:0000000338945889)↗

Robust analysis of space-, time-, and energy-resolved soft x-ray measurements of magnetically confined fusion plasmas (invited)

A novel compact multi-energy soft x-ray (ME-SXR) diagnostic based on the PILATUS3 100K x-ray detector has been developed in collaboration between the Princeton Plasma Physics Laboratory and the University of Wisconsin-Madison and tested on the Madison Symmetric Torus (MST) reversed-field pinch. This solid-state photon-counting detector consists of a two-dimensional array of ~100 000 pixels for which the lower photon absorption cutoff energy can be independently set, allowing it to be configured for a unique combination of simultaneous spatial, spectral, and temporal resolution of ~1 cm, 100 eV, and 500 Hz, respectively. The diagnostic is highly versatile and can be readily adapted to diverse plasma operating conditions and scientific needs without any required downtime. New results from improved-confinement and quasi-single helicity plasmas in the MST demonstrate how the detector can be applied to study multiple aspects of the evolution of magnetically confined fusion-grade plasmas. These include observing the evolution of thermal emissivity, characterizing the energy of mid-Z excitation lines, extracting the T e profile, and observing the evolution of non-thermal populations. A technique for integrating the ME-SXR diagnostic into an integrated data analysis framework based on Bayesian inference is also presented. As a result, this allows ME-SXR measurements to be combined with data for complementary diagnostics in order to simultaneously infer Te and nZ from all available information.

47 OTHER INSTRUMENTATION↗

Using integrated data analysis to extend measurement capability (invited)

The analysis approach called integrated data analysis (IDA) provides a means to exploit all information present in multiple streams of raw data to produce the best inference of a plasma parameter. This contrasts with the typical approach in which information (data) from a single diagnostic is used to measure a given parameter, e.g., visible bremsstrahlung → Z eff . Data from a given diagnostic usually contain information on many parameters. For example, a Thomson scattering diagnostic is sensitive to bremsstrahlung and line emission in addition to electron temperature. This background light is typically subtracted off and discarded but could be used to improve knowledge of Z eff . IDA encourages explicit awareness of such information and provides the quantitative framework to exploit it. This gives IDA the ability to increase spatial and temporal resolution, increase precision and accuracy of inferences, and measure plasma parameters that are difficult or impossible to measure using single diagnostic techniques. One example is the measurement of Z eff on Madison symmetric torus using IDA since no single diagnostic can provide a robust measurement. Thus, as we enter the burning plasma era, application of IDA will be critical to the measurement of certain parameters, as diagnostic access in the harsh fusion environment will be extremely limited.

70 PLASMA PHYSICS AND FUSION TECHNOLOGY↗

Incorporating Beam Attenuation Calculations into an Integrated Data Analysis Model for Ion Effective Charge

Here, we have created a forward model for charge-exchange impurity density measurements that incorporates neutral beam attenuation measurements self-consistently for the purpose of determining the ion-effective charge Z eff . The model is constructed within an integrated data analysis framework to include a self-consistent calculation of neutral beam attenuation due to multiple impurity species into the measurement of a single impurity density. The model includes measurements of the beam Doppler-shift spectrum and shine-through particle flux to determine the neutral beam particle density which is attenuated by ion collisions. Synthetic data are generated from the diagnostic forward model using statistical and calibration uncertainties. These “noisy” data are used in the analysis to evaluate how accurately Z eff is determined. Methods of experimental design are employed to calculate the information gained from different diagnostic combinations. The analysis shows that while attenuation measurements alone do not provide a unique impurity density measurement in the case of a multispecies inhomogeneous plasma, they do provide an effective measurement of the Z eff profile and place constraints on the impurity density profiles.

70 PLASMA PHYSICS AND FUSION TECHNOLOGY↗

A comparison between soft x-ray and magnetic phase data on the Madison symmetric torus

The Soft X-Ray (SXR) tomography system on the Madison Symmetric Torus uses four cameras to determine the emissivity structure of the plasma. This structure should directly correspond to the structure of the magnetic field; however, there is an apparent phase difference between the emissivity reconstructions and magnetic field reconstructions when using a cylindrical approximation. The difference between the phase of the dominant rotating helical mode of the magnetic field and the motion of the brightest line of sight for each SXR camera is dependent on both the camera viewing angle and the plasma conditions. Holding these parameters fixed, this phase difference is shown to be consistent over multiple measurements when only toroidal or poloidal magnetic field components are considered. These differences emerge from physical effects of the toroidal geometry which are not captured in the cylindrical approximation.