Thermal-Expansion Measurement
Precise stable laser system determines coefficients of thermal expansion. Dual-beam interferometer arrangement monitors changes in sample length as function of temperature by following changes inoptical path lengths.
Engineering topics
Publications and source records attributed to Rives, C..
Precise stable laser system determines coefficients of thermal expansion. Dual-beam interferometer arrangement monitors changes in sample length as function of temperature by following changes inoptical path lengths.
A system for measuring the expansion of low coefficient of thermal expansion (CTE) materials was constructed around a H.P. 5526-A laser measuring system. The vacuum CTE measurements in the -150 F to +120 F range were made over a 6 month period on a graphite epoxy tube yielding CTE values of 2.5 to one fifty-millionth/F above ambient and 2 + or - one ten-millionth F below ambient temperature. To assure that the below ambient, approximately 10 microns high open loop nature of the delta L/L vs. T curves was not apparatus related, similar size quartz tubes (A and B) were checked and found to have only a 2 micron (negligable for quartz) open loop component. These two quartz tubes, A and B, had ambient CTE values 20% and 45% respectively higher than the average handbook value. The overnight microcreep diminished an order of magnitude during the first several cycles after the system had been reopened.