Search NASA⌕ Search

Engineering topics

Sadleir, John E

Publications and source records attributed to Sadleir, John E.

Longitudinal Proximity Effects in Superconducting Transition-Edge Sensors

We have found experimentally that the critical current of a square thin-film superconducting transition-edge sensor (TES) depends exponentially upon the side length L and the square root of the temperature T, a behavior that has a natural theoretical explanation in terms of longitudinal proximity effects if the TES is regarded as a weak link between superconducting leads. As a consequence, the effective transition temperature T{sub c} of the TES is current dependent and at fixed current scales as 1/L{sup 2}. We have also found that the critical current can show clear Fraunhofer-like oscillations in an applied magnetic field, similar to those found in Josephson junctions. We have observed the longitudinal proximity effect in these devices over extraordinarily long lengths up to 290 {micro}m, 1450 times the mean-free path.

75 CONDENSED MATTER PHYSICS, SUPERCONDUCTIVITY AND↗

Optimizing Transition-Edge Sensor Design for High Count-Rate Applications

We are developing transition-edge sensor (TES) X-ray detectors optimized for high count-rate applications. These devices are fabricated on thick (300 {mu}m) Si substrates, resulting in a 20 times increase in thermal conductance to the heat sink compared to our conventional membrane isolated TES's. Operating a TES with higher heat sink conductance requires 4.5 times more bias current. This results in a 2.7 times increase in {beta}, the logarithmic derivative of resistance with respect to current. Noise measurements show a lower limit on the TES excess noise scales as (2{beta}){sup 1/2}, consistent with the near-equilibrium, non-linear expansion of the Ohmic Johnson noise. This is consistent with our membrane devices though the increased {beta} means the theoretical best attainable resolution is degraded by 25-35%. We have tested devices with different contact geometries between the absorber, and the TES and substrate. This allows us to investigate the loss of athermal phonons to the substrate, which can degrade the resolution. Results show a correlation between the stem contact area and a low-energy tail in the spectral response at 5.9 keV due to the athermal phonon loss. In several devices tested we demonstrate a resolution of 4.1-5.6 eV, coupled with detector time constants as fast as 44 {mu}s, representing an increase in detector response by 7 times compared to the membrane devices.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Experimental Results and Modeling of Low-Heat-Capacity TES Microcalorimeters for Soft-X-ray Spectroscopy

Transition-edge-sensor (TES) X-ray microcalorimeters have mostly been targeted at mid-band energies from 0.05-10 keV and high energies to above 100 keV. However, many other optimizations are possible. Here we present results from devices optimized for soft X-ray applications. For spectroscopy below 1 keV, the X-ray stopping power and heat capacity (C) of the TES itself are high enough that we can omit a separate absorber. The resulting devices have low C and the best-achievable energy resolution should be under 1 eV. We are interested in pursuing such devices primarily for astrophysical applications and laboratory astrophysics at LLNL's Electron-Beam Ion Trap. To this end, we have studied arrays in which 'bare' TESs are interspersed with broad-band pixels that have absorbers. By extending the absorbers to cover the area where the leads contact the low-energy pixels, we have eliminated a significant source of non-Gaussian detector response. The bare devices are in a different regime from our typical devices in that C is ten times lower and the conductance to the bath is four times lower. We have explored this regime through simultaneous fitting of noise and impedance data. These data cannot be fit by the simple model we employ to describe our typical broad-band devices. In this contribution we present X-ray spectra and the results from modeling.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗