Application of Tunable Diode Laser Spectrometry for Isotopic Analysis

Lead-salt tunable diode lasers are now commercially available with operating temperatures above 77 K and excellent single-mode characteristics which make them very useful for spectroscopy in the mid-IR spectral range. Using these lasers for absorption measurements in the 4.3 micrometer spectral region we have made accurate measurements of isotopic abundances in CO2. The method involves assembly language controlled data acquisition of spectra of selected adjacent isotopic rotational lines, an etalon spectral frequency calibration technique, the fitting of multiple Voigt profiles to the data, and the ratioing of absorbances from simultaneously acquired sample and reference data scans. The accuracy achieved is better than 0.4% for the C-13/C-12 ratio and 1% for the O-18/O-16 ratio. Higher accuracy is expected from improvements now being implemented.

Sauke, T. B.↗

Stable carbon isotope measurements using laser spectroscopy

The 2300 cm(exp -1) spectral region is especially interesting because (12)CO2 and (13)CO2 bands overlap in such a way that their rotational lines have approximately equal absorbance at the anticipated isotopic ratio (approximately 90) of carbon on Earth and Mars. Pairs of rotational lines we have studied are separated by as little as 0.050 cm(exp -1), but are well resolved with a tunable diode laser. Using sophisticated sweep integration and signal averaging techniques, we have measured the stable isotope ratio in carbon dioxide to a precision of better than 1 percent.