Variations in in-flight absolute radiometric calibration
Variations in the in-flight absolute radiometric calibration of the Coastal Zone Color Scanner and the Thematic Mapper (TM) are reviewed. At short wavelengths, the sensors show a gradual reduction in response, while in the mid-IR the TM shows oscillatory variations. One set of measurements made at White Sands, New Mexico shows anomalous results in TM bands 2 and 4. The results of a reflectance-based and a radiance-based calibration method at White Sands are described. An analysis of the radiance-based method shows the value of such measurements from helicopter altitudes for calibration.
Slater, Philip N.↗