Search NASA⌕ Search

Engineering topics

Smith, Susan K.

Publications and source records attributed to Smith, Susan K..

Summary of the Workshop on M400 High-Resolution CZT Detector Safeguards Applications

In September 2023, the US Department of Energy (DOE) National Nuclear Security Administration’s (NNSA’s) Office of Nonproliferation and Arms Control sponsored a workshop on M400 high-resolution cadmium zinc telluride (CZT) detector safeguards applications under its Safeguards Technology Development Program. The workshop brought together internationally recognized subject matter experts (SMEs) in the field of gamma spectrometry and uranium/plutonium isotopic analysis and experts from H3D Inc. (H3D), the vendor for M400.

98 NUCLEAR DISARMAMENT, SAFEGUARDS, AND PHYSICAL P↗

Measurement Plan for Uncertainty Contributions to 252 Cf Waste Measurements

This measurement plan establishes the methodology for performing and analyzing mock-up measurements of Building 7930 Cell G Cf waste generated from 252 Cf product preparation to better determine the measurement uncertainty contributors. Understanding the uncertainty contributors will establish the total measurement uncertainty. This is critical because it will impact the administrative threshold at which personnel can discriminate between low-level waste (LLW) and transuranic (TRU) waste.

12 MANAGEMENT OF RADIOACTIVE AND NON-RADIOACTIVE W↗

The Impact of Increased Requirements for Verification Measurements of Category III and IV Receipts

As part of the regular nuclear material accounting and control process, sites that have accountable nuclear material regularly perform measurements of the material in their possession. The expectation for the type of measurements and their frequency is set out in Department of Energy (DOE) Order 474.2, and further detailed in each site’s respective material control and accountability (MC&A) plan and associated controlling documents. The measurements of interest in this report are those performed on external (between DOE sites) transfer receipts of Category III or IV accountable materials. In 2020, a revision of DOE Order 474.2 was proposed that would significantly increase the requirement that measurements be performed upon receipt of material, and further, that those measurements be quantitative verification (rather than qualitative confirmatory) measurements. A survey was performed of DOE sites to discuss the impact on their MC&A measurements program if more rigorous verification measurements requirements for Category III and IV receipts were implemented. An assessment of nondestructive assay (NDA) measurement instruments to meet those needs was also performed. Across the board, sites indicated that personnel and budget would cause the largest impact. Sites which are qualified as Category I/II indicated a reduced impact as these sites already have the measurement capabilities and some personnel but would require the acquisition of additional measurement systems. Category III/IV sites indicated a much larger impact as these sites generally have smaller MC&A programs. A few smaller sites stated a measurement program would need to be created to handle the demand. Calorimetry, quantitative gamma spectroscopy and neutron coincidence counting systems were identified as common systems that could meet the verification measurement requirements. The other, and perhaps larger issue is the need for calibration standards for the calorimetry and neutron coincidence counting systems. It is proposed that a comparative study be performed for ISOCS and ISOTOPICS, qualification of a COTS calorimetry system be performed, and development of an AWCC system that uses a neutron generator.

98 NUCLEAR DISARMAMENT, SAFEGUARDS, AND PHYSICAL P↗

Analysis of FRAM Measurements to Establish Methodology Uncertainty

Qualification measurements were performed at Oak Ridge National Laboratory (ORNL) on a variety of measurement standards. These measurements will be used to qualify the Fixed-Energy, Response Function Analysis with Multiple Efficiency (FRAM) code with a high purity germanium (HPGe) detector for use at ORNL. Measured items include uranium isotopic standards, U 3 O 8 line sources, and a uranium tetrafluoride source. The 235 U enrichment was estimated using the FRAM isotopic analysis software for each source and compared with the declared isotopic values.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Analysis of ISOCS Measurements to Establish Methodology Uncertainty

Qualification measurements were performed at the US Department of Energy’s (DOE’s) Oak Ridge National Laboratory (ORNL) on a variety of measurement standards. These measurements will be used to qualify the In-Situ Object Counting System (ISOCS) measurement method with a high-purity germanium (HPGe) detector for use at ORNL. Measured items include uranium isotopic standards, filter papers coated with uranium oxide (U 3 O 8 ), U 3 O 8 line sources, and a uranium tetrafluoride (UF 4 ) source. The 235 U mass was estimated using the ISOCS calibration software for each source and compared with the declared isotope mass value.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Cell G 7606A Mockup Measurement Analysis Methods and Results

The production of research isotopes at the Radiochemical Engineering Development Center (REDC) invariably leads to the accumulation of radioactive waste. Storage of this waste onsite at ORNL serves as an interim solution prior to shipment to long term storage facilities, however onsite capacity is limited. Proper characterization of the activity of waste products is essential for determining the appropriate waste stream and, ultimately, mitigating the cost of disposal. This is typically done via gamma spectrometry and use of the In-Situ Object Counting System (ISOCS), a software package from Mirion that serves as an accepted community standard. However, variability in the contents, density, and activity distribution in waste containers can introduce large errors in ISOCS quantification. A measurement campaign using a mockup of the proposed setup in Cell G seeks to quantify the magnitude and source of these systematic errors. Analysis of this data will guide the creation of ISOCS geometry templates and measurement methods designed to minimize overall uncertainty in reported activities.

12 MANAGEMENT OF RADIOACTIVE AND NON-RADIOACTIVE W↗

Thermal impacts on nondestructive analysis measurements of uranium hexafluoride

A series of nondestructive assay measurements were performed before, during, and after localized heating of an industrial standard 12B cylinder of uranium hexafluoride to replicate dose variations previously observed on commercial 30B cylinders. While cylinders represent a closed system, enrichment measurements may be impacted by altered uranium and daughter product distributions due to sublimation. A series of measurements were performed to examine localized heating impacts on dose rates, ratios of uranium and daughter product distributions, and measured 235 U enrichment as calculated by two different approaches. The less complex enrichment meter approaches were found to be more robust and less prone to deviation than isotope ratio approaches such as FRAM, but the impact can be minimized by using observable dose changes to avoid impacted areas.

38 RADIATION CHEMISTRY, RADIOCHEMISTRY, AND NUCLEA↗

Heterogeneity effects on nondestructive assay measurements of enrichment in UF 6 cylinders

A set of measurements were performed using multiple mechanically cooled HPGe detectors at six positions around a cylinder uranium hexafluoride to assess if matrix inhomogeneity is detectible and effects the determined apparent uranium enrichment. These measurements were performed on nine 30B cylinders of uranium hexafluoride. Six of the cylinders appeared to be homogeneous and had similar apparent measured uranium enrichment, as calculated by the FRAM software, at all positions. However, three cylinders appeared to have local inhomogeneity based on the observed results. This was manifested as very low apparent enrichments, often <10% of the declared enrichment, on select positions. The side with a low apparent enrichment shows elevated daughter isotope activity (e.g., elevated 1001 keV and 766 keV peak count rate) and slightly reduced 235 U activity (lower 185.7 keV count rate). We hypothesize that these heterogenous cylinders may have experienced asymmetric solar heating, which caused volatile UF 6 to sublime preferentially away from the warmed side. A similar phenomenon was observed at a second facility where dose rate measurements confirm that the “low-enriched side” of cylinders show elevated gamma ray dose rates, likely from the removal of UF 6 attenuation and concentration of daughter-products. Care should be taken during uranium enrichment verification when applying methods that include gamma-rays associated with daughter nuclides to the determination of uranium enrichment on cylinders that are stored in sunlight and have asymmetric dose readings around the cylinder.

11 NUCLEAR FUEL CYCLE AND FUEL MATERIALS↗

Geology and Design of Major Spent Fuel Repositories

This report summarizes the geology and design concepts of four major spent fuel repositories from around the world. The crystalline rock with bentonite backfill is currently implemented in Finland with another repository being constructed in Sweden. The salt formation design focuses on the Waste Isolation Pilot Plant (WIPP) repository in the United States. The volcanic tuff open air design is based on the Yucca Mountain location. The volcanic tuff open air design at Yucca Mountain is only a concept as this repository has not been approved for operations. The last spent fuel repository design, clay/shale formation with backfill, focuses on the Cigeo Project in France, which is also pending.

12 MANAGEMENT OF RADIOACTIVE AND NON-RADIOACTIVE W↗

Comparison of gamma-ray spectral analysis methods for thick-walled UF 6 cylinders

Three data analysis methods were compared for gamma ray spectra collected by planar and coaxial high-purity germanium detectors (HPGe) on thick-walled uranium hexafluoride cylinders. The cylinder contents range from natural to low-enriched uranium (4.95% 235 U). Each spectrum was analyzed with the traditional 185.7 keV enrichment meter method, a 185.7/1001 keV ratio method, and the Fixed Energy Response Function Analysis with Multiple Efficiency (FRAM) software that utilizes low- and high-energy gamma rays and X-rays. When results for all cylinders were compared, the enrichment meter method and the 185.7/1001 keV peak ratio method consistently provided the most accurate data for all detector types. The error on the enrichment calculations analyzed with FRAM did not meet the international target value whereas those calculated with the enrichment meter and ratio method did. Enrichments calculated by FRAM were found to skew to higher values on natural uranium and to lower values on low-enriched uranium. FRAM calculated the least-precise enrichments; replicate analyses of single cylinders resulted in 235 U values that varied by as much as 19%. FRAM does have the significant advantages in that it does not require the use of calibration cylinders, infinite thickness criteria are not necessary, and it can be applied to any HPGe detector. Finally, if calibration cylinders are available, the enrichment meter and 185.7/1001 peak ratio are the recommended methods as they are the most accurate and precise.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Applications of HPGe-detected high energy gamma rays toward quantifying neutron emission rates and 234 U enrichment in UF 6 cylinders

This study presents a 234 U enrichment meter that uses high energy gamma and neutron count rate as measured by a portable, high-purity germanium gamma ray detector. Twenty-one 30B cylinders containing UF 6 of varying origin were analyzed with an HPGe detector for high-energy gamma ray and neutron counts. Excellent correlations (R 2 0.95) between various regions of 3.0 MeV gamma rays, neutron count rate, and uranium enrichment were found. These neutron and high-energy gamma data are used to create calibration-dependent enrichment meters for 234 U. The 234 U enrichment calculated from the neutron count rate and 4 – 6 MeV gamma count rate is more accurate than methods that rely on the 120.9 keV peak which is strongly attenuated inside commercial sized UF 6 cylinders. These neutron and high-energy gamma enrichment calculators can also be used to determine a 235 U enrichment, assuming mass-based enrichment from natural uranium, but the results are not as accurate as the traditional 185.7 keV peak enrichment meter.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗