Swab Testing Results

1. All 3D printed swabs bend to at least 90° without breaking. 2. All 3D printed swabs are at least as strong under tensile testing as their commercial counterparts. a. The Abiogenix FAST Spiral NP and EnvisionTEC swabs retain their tensile strength after sterilization and after 7 days of aging. b. For the Formlabs swabs, steam sterilization significantly reduces their tensile strength, which mostly recovers after 7 days of aging. However, the swabs are stronger than commercial ones at all times. c. The Resolution Medical Lattice swabs retain their tensile strength after 5 days of aging. Five different swab types were tested: 1. Abiogenix FAST Spiral NP Swab; 2. Formlabs NP Swab; 3. EnvisionTEC NP Swab; 4. Resolution Medical Lattice Swab; and 5. Fosun Pharma NP Swab The Abiogenix FAST Spiral NP, EnvisionTEC NP, Resolution Medical Lattice, and Fosun Pharma NP swabs were provided by their respective suppliers. The Formlabs NP swabs were printed at LLNL using the Form 3B Printer and Formlabs Surgical Guide resin

59 BASIC BIOLOGICAL SCIENCES↗

Swab Testing Results Summary

Swabs Tested: 1. FAST Spiral NP Swab, designed and supplied by Abiogenix, printed from polyamide 11 (Nylon) using HP Multi Jet Fusion printers. 2. NP Swab designed by USF Health and Northwell Health in collaboration with Formlabs. Printed at LLNL using the Form 3B stereolithography printer in Formlabs Surgical Guide resin.

42 ENGINEERING↗

Swab Tensile Testing Results and Procedures - PA11 Rev. 10 Swabs

This tensile test is to mimic the worst-case scenario of the swab being caught on an obstruction when being pulled out of the nasopharyngeal space. It is to determine how much tensile force the swab can withstand without breaking. All swabs were provided by HP and were the PA11 material and Rev 10 design. All of the swabs were tested using the protocol outlined in Section 2 (Tensile Testing Procedure). Not all of the swabs were pre-treated using the Autoclave. For those swabs that were pre-treated using the Autoclave, the tensile tests were performed within 6 hours of completion of the Autoclave pre-treatment, with the exception of the “Aged” swabs. For the “Aged” swabs, the tensile tests were performed 7 days after the Autoclave pre -treatment. The “Aged” swabs that were not pre-treated using the Autoclave, were tested 7 days after those that were not “Aged.”

36 MATERIALS SCIENCE↗

Swab Tensile Testing Results and Procedures

This tensile test is to mimic the worst-case scenario of the swab being caught on an obstruction when being pulled out of the nasopharyngeal space. It is to determine how much tensile force the swab can withstand without breaking. A variety of different swabs (different materials and different geometries) were tested using the protocol outlined in Section 2 (Tensile Testing Procedure). Not all of the swabs were pre-treated using the Autoclave. For those swabs that were pre-treated using the Autoclave, the tensile tests were performed within 6 hours of completion of the Autoclave pre-treatment, with the exception of the “Aged” swabs. For the “Aged” swabs, the tensile tests were performed 7 days after the Autoclave pre -treatment. All swabs were provided by HP.

36 MATERIALS SCIENCE↗

Swab Tensile Testing - HP Rev 10 Summary

This tensile test is to mimic the worst-case scenario of the swab being caught on an obstruction when being pulled out of the nasopharyngeal space. It is to determine how much tensile force the swab can withstand without breaking.

42 ENGINEERING↗

Swab Testing Results- HP PA11 Rev 8 Summary

This tensile test is to mimic the worst-case scenario of the swab being caught on an obstruction when being pulled out of the nasopharyngeal space. It is to determine how much tensile force the swab can withstand without breaking.

42 ENGINEERING↗

Swab Tensile Testing and Procedures

This tensile test is to mimic the worst-case scenario of the swab being caught on an obstruction when being pulled out of the nasopharyngeal space. It is to determine how much tensile force the swab can withstand without breaking. A variety of different swabs (different materials and different geometries) were tested using the protocol outlined in Section 2 (Tensile Testing Procedure). Not all of the swabs were pre-treated using the Autoclave. For those swabs that were pre-treated using the Autoclave, the tensile tests were performed within 6 hours of completion of the Autoclave pre-treatment, unless otherwise noted. The FormLabs-USF and FormLabs-Northwell were printed at LLNL using FormLabs Surgical Grade V1 Resin on a Form 3B Printer. Results Summary: All FormLabs design and treatment variations (e.g. autoclaving and “aging”) had higher tensile strength compared commercial swabs. Autoclaving degraded the tensile strength. “Aging” after autoclaving improved the tensile strength. “Aging” without autoclaving degraded the tensile strength. No differences were observed in tensile strength with different autoclaving protocols. No significant differences were observed in tensile strength for different batches.

59 BASIC BIOLOGICAL SCIENCES↗

Swab Tensile Testing Results [Slides]

HP PA11 material swabs predominantly broke at the joint between the head and the neck, although there was significant variation in the break locations. HP PA12, Injection Molded, and FormLabs (Surgical Guide V1 Resin) material swabs predominately broke in the neck region.

36 MATERIALS SCIENCE↗

Swab Tensile Testing Results and Procedures

This tensile test is to mimic the worst-case scenario of the swab being caught on an obstruction when being pulled out of the nasopharyngeal space. It is to determine how much tensile force the swab can withstand without breaking. A variety of different swabs (different materials and different geometries) were tested using the protocol outlined in Section 2 (Tensile Testing Procedure). Not all of the swabs were pre-treated using the Autoclave. For those swabs that were pre-treated using the Autoclave, the tensile tests were performed within 6 hours of completion of the Autoclave pre-treatment, with the exception of the Injection Molded samples. For the Injection Molded samples, the tensile tests were performed within 24 hours of the completion of the Autoclave pre-treatment. The FormLabs-USF and FormLabs-Northwell were printed at LLNL using FormLabs Surgical Grade V1 Resin on a Form 3B Printer. All other samples were provided by HP.

36 MATERIALS SCIENCE↗