Photoneutron Detection in Active Interrogation Scenarios Using Small Organic Scintillators
Not provided.
Engineering topics
Publications and source records attributed to Steinberger, William M..
Not provided.
Organic glass scintillator (OGS) bars (6 × 6 × 50 mm 3 ) were melt-cast and characterized with dual-ended readout by silicon photomultiplier arrays. The results were compared to trans-stilbene bars of the same size. Energy resolution was measured via a 137 Cs Compton-coincidence experiment. Time resolution was determined via a 22 Na coincidence-timing experiment. Depth-of-interaction resolution was measured by moving a gamma ray fan beam along the major axis of the bars. Pulse shape discrimination capability and neutron light output were determined via a time-of-flight measurement of a 252 Cf source. The OGS bars exhibited better energy resolution at 478 keV (10.3 ± 0.5% vs. 11.2 ± 0.5%), better time resolution from 200–341 keVee (270 ps vs. 340 ps FWHM), slightly poorer depth-of-interaction resolution, poorer PSD performance, and higher neutron light output than the stilbene bars. Here, the low cost, ease of manufacturing, brightness, and excellent time resolution make OGS a strong candidate for use in compact radiation imaging systems.
Not Available