Hubble space telescope component qualification utilizing a residual gas analyzer
An approach is described for integrating an automated residual gas analyzer (RGA) into the thermal vacuum outgas process. The RGA has the capability of detailed analysis over a 200 amu range with monitoring of each peak as the test progresses. The NASA criterion of either a 50% reduction of each peak over 44 amu or less than an 0.1% amplitude of the 28 amu base peak for all amu's above 44 amu is satisfied. A computer sorts, reports which peaks do not satisfy the criterion, and reports test completion. The computer plots amplitudes of all the peaks from 1 amu through 200 amu for each scan. This RGA real time approach is unique in that it provides sufficient detail to highlight a problem as it occurs. By comparison, a quartz crystal microbalance (QCM) or a thermoelectric quartz crystal microbalance (TQCM) indicates to the operator when changes occur, but not what has happened. Optical witness samples indicate whether a test was satisfactory within a few days after the test was completed.