The continuous measurement of hydrogen chloride in the ambient atmosphere using the dual isotope infrared absorption technique
The results of a program to develop a prototype gas filter correlation NDIR analyzer capable of providing the required HCl measurement capability, while maintaining an adequate rejection of any other gases anticipated in the atmosphere are presented. Examples of the performance of the prototype analyzer are presented which show an rms noise equivalent concentration of 0.06 ppm of HCl was achieved while maintaining an electronically determined 10% to 90% time response to gas samples of about 2 seconds. No measureable response was observed to CO2, CO, and H2O while maintaining an adequate rejection of the hydrocarbons, for example CH4 and n-hexane. The experiments were performed which demonstrate that the span stability of the HCl gas filter correlation analyzer is unaffected by the presence of water vapor and which support the belief that the incorporation of a relatively open-volume, multiple path sample cell into the instrument would enable ground station as well as airborne measurements of trace quantities of HCl in the ambient atmosphere to be performed.