MODIS Reflective Solar Bands Calibration Improvements for Collection 7

Calibration of Terra and Aqua MODIS reflective solar bands (RSB) has evolved significantly since the launch of the first MODIS instrument on the Terra satellite more than 21 years ago. In NASA’s current Collection 6 and 6.1 Level 1B products (C6/C6.1 L1B), the RSB calibration algorithm continues to rely primarily on the onboard solar diffuser to calibrate the instrument gain. Lunar observations are used to track on-orbit changes in the response versus scan angle (RVS), and data from pseudo-invariant desert sites are used to apply adjustments to the gain and RVS calibration for select bands. The resulting reflectance products have in general shown a very stable performance. In recent years, some performance degradation has been noted for a few bands and algorithm changes have been tested to further improve the calibration accuracy for the upcoming Collection 7 (C7) L1B reprocess. In this paper, we present the MODIS RSB calibration improvements that will be included in C7. Major improvements include: applying polarization correction to the desert data before using it to generateRVS for Terra bands 8, 9, 3, and 10; using ocean scene data and an interband calibration approach to correct for long-term drift of Terra bands 11 and 12; applying an updated crosstalk correction to Terra SWIR bands over the entire mission; and using data from deep convective clouds in Terra SWIR band calibration, including the addition of time-dependent RVS for bands 5 and 26. All other minor calibration changes are also covered. Overall, the reflectance differences at nadir between C6.1 and C7 are within a few percent, though the differences increase in some cases at large scan angles. The Terra visible (3, 8-12) and SWIR bands (5-7, 26) have the most significant improvements. For all other Terra bands and all Aqua bands, the C7-C6.1 differences are mostly within 1%

MODIS

MODIS Thermal Emissive Bands Calibration Improvements for Collection 7

The MODIS thermal emissive bands (TEB) radiometric calibration uses a quadratic function for the instrument response, and the calibration coefficients look-up tables (LUTs) are updated using the response of an on-board blackbody (BB). After more than 21 and 19 years on-orbit, the TEB performances for Terra and Aqua MODIS have been generally stable. However, contamination from electronic crosstalk, a known issue since prelaunch, has affected the L1B image quality and measurement accuracy. In addition to the photovoltaic (PV) longwave infrared (LWIR) bands crosstalk correction included in Terra MODIS Collection 6.1 (C6.1), a crosstalk correction for select detectors in the Terra and Aqua mid-wave infrared (MWIR) and Aqua PV LWIR bands are applied in C7. The mission-long crosstalk coefficients for the selected detectors are derived and populated in the form of LUTs. The crosstalk correction is applied to both on-orbit calibration and the algorithm used to generate Earth-view L1B products. Among these detectors, the Aqua MODIS band 24 detector 1 crosstalk has the largest impact on image quality, with striping observed over cold scenes for both Terra and Aqua MODIS. The images of C7 L1B and C6.1 are compared to assess the impact of the correction. Additional assessments using Earthview measurements and inter-comparison also revealed the need for improvement of calibration stability and consistence for select bands. Additional improvements for long-term stability and mirror side consistence were developed using quasideep convective clouds (qDCC), Dome-C, ocean, desert, and inter-comparison with other instruments.

MODIS

On-Orbit Calibration and Performance Assessments of Terra and Aqua Modis Thermal Emissive Bands

Terra and Aqua MODIS have successfully operated for more than 20 and 18 years, respectively, and far exceeded their designed lifetimes of 6 years. MODIS has 36 spectral bands, among which 16 are the thermal emissive bands (TEB) covering a wavelength range from 3.75 to 14.24 μm. Observations from the MODIS TEB have been used to generate a number of data products, such as surface/cloud/atmospheric temperatures, cloud top altitude, and water vapor properties. To a large extent, the quality of the MODIS L1B radiance product has been maintained through the entirety of each mission via extensive on-orbit calibration and validation efforts. MODIS TEB calibration uses a quadratic algorithm and is referenced to an on-board blackbody with its temperature measurements traceable to the NIST standard. We provide an overview of the MODIS instrument operations, key TEB calibration activities, and algorithms used in the latest L1B data Collection 6.1 (C6.1) and describe the TEB on-orbit performance for both Terra and Aqua MODIS. The TEB gain trends have been stable to within 5% over the Terra mission after 2003 under the same electronic configuration, except for bands 27 to 30, and within 3% over the Aqua mission. The Terra MODIS optical crosstalk correction implemented right after launch has been very effective in removing the ghost images in its photoconductive detector bands 32 to 36. Starting from C6.1, an electronic crosstalk correction algorithm has also been applied to the Terra MODIS photovoltaic detector long-wave infrared bands 27 to 30. The noise characterization performance remains stable with most TEB detectors continuously meeting their design requirements. A gradual loss of the Aqua MODIS radiative cooler margin reached its maximum in 2013 and has been slowly recovering since then. On-orbit changes in TEB response versus scan angle are extremely small based on pitch maneuver observations and assessments using vicarious approaches. The remaining challenges in TEB calibration and proposed improvements for the upcoming data collection with L1B data production and reprocessing are discussed.

Terra

Improvements of On-orbit Characterization of Terra MODIS Short-wave Infrared Spectral Bands Out-of-Band Responses

The short-wave infrared (SWIR) bands (5-7, 26) of Terra MODIS, which are co-located with the mid-wav e infrared (MWIR) bands (20-25) on the short and mid-wave infrared (SMIR) Focal Plane Assembly (FPA),have a known issue related to 5.3 μm out-of-band (OOB) thermal leak and electronic crosstalk that was identified prelaunch. As a result, a crosstalk correction algorithm was designed and implemented in the MODIS Level 1B (L1B) calibration. Shortly after the Terra launch, extensive efforts were under taken to help characterize and mitigate the impact due to the OOB response and crosstalk on the SWIR on-orbit calibration and, consequently, the associated L1Bdata products. In addition, special night time day mode (NTDM) operations have been regularly scheduled to derive the crosstalk correction coefficients. Since MODIS does not have a spectral band centered at 5.3 μm, its band 28 (7.325 μm) was chosen as the surrogate sending band to simulate the OOB radiances at 5.3 μm. This was largely based on the measurements from the MODIS Airborne Simulator (MAS) spectrometer field campaigns in the early months after the Terra launch. In the case of Aqua MODIS, the magnitude of the SWIR crosstalk was much smaller and band 25 (4.52 μm) was found to be more effective as the sending band for the crosstalk correction. In recent years, the Terra MODIS photovoltaic (PV) long-wave infrared (LWIR) bands (27-30)electronic crosstalk has increased considerably, especially after the spacecraft safe-mode event occurred in February2016. This accentuated degradation in the PV LWIR performance has also impacted the performance of the SWIR crosstalk correction and thus its calibration and data quality. In this paper, we examine the use of band 25 as the sending band for the Terra MODIS SWIR crosstalk correction and compare its performance with that based on band 28 as the sending band. Results indicate an improvement of on-orbit gain stability for the SWIR calibration and reduced detector to detector and sub-frame to sub-frame striping in the calibrated L1B imagery, especially during the period when the PV LWIR electronic crosstalk has become more severe. This approach has been implemented in the forward production of Terra MODIS Collection 6 and Collection 6.1,starting from July 2019, and is planned to be used for the future reprocessing of MODIS L1B and to help improve the mission-long reflectance calibration and trending stability of the SWIR bands.

MODIS, Terra, Aqua, MODIS, SWIR, calibration, out-

The MODIS RSB calibration and look-up-table delivery process for Collection 6 and 6.1

The Moderate Resolution Imaging Spectroradiometer (MODIS) instruments on-board the Aqua and Terra space-craft have provided valuable science data for the last 18 and 20 years, respectively. Each instrument is equipped with 36 spectral bands, 20 of which are reflective solar bands (RSBs). These bands cover a wavelength range of 0.4 – 2.2μm and are calibrated on-orbit using several on-board calibrators (OBCs), such as a solar diffuser (SD)and a solar diffuser stability monitor (SDSM), along with regularly-scheduled lunar observations through the space view (SV) port. The gain (1/m1) and response-verses-scan angle (RVS) are updated on a near-monthly basis and act as the primary look-up-tables (LUTs) for the RSB calibration. A set of separate uncertainty LUTs for each of the RSBs are also delivered regularly and incorporated into the Level 1B (L1B) product to generate a pixel-level Uncertainty Index (UI). In addition to the gain, RVS and uncertainty, there are several other LUTs associated with the reflective bands that are either updated less frequently or remain static. The accuracy of both the forward-predicted and historical RSB LUTs, which are derived by the MODIS Characterization Sup-port Team (MCST), is important in maintaining the quality and accuracy of the L1B and science products.To ensure a timely and accurate LUT update, MCST has established a comprehensive set of procedures. This paper provides an overview of the calibration process, along with the current LUT delivery process for the RSB sin Collection 6 (C6) and Collection 6.1 (C6.1). Improvements to be implemented in future collections are also discussed.

MODIS

MODIS and VIIRS Calibration and Characterization in Support of Producing Long-Term High-Quality Data Products

Terra and Aqua Moderate Resolution Imaging Spectroradiometer (MODIS) have successfully operated since their launches in 1999 and 2002, respectively, and generated various data products to support the Earth remote sensing disciplines and users worldwide for their research activities and applications, including studies of the Earth system, and its changes over time and geographic regions. The MODIS data have also significantly contributed to the continuity of multi-decadal satellite data records and led to major advances in the Earth remote sensing field. The long-term data records from MODIS observations have been and will continue to be extended by the Visible Infrared Imaging Radiometer Suite (VIIRS) instruments, currently operated aboard the Suomi-National Polar-Orbiting Partnership (NPP) and NOAA-20 satellites. The data quality of satellite instruments strongly depends on their calibration accuracy and stability. In order to help scientists and users gain a better understanding of MODIS and VIIRS data quality, this paper provides an overview of their on-orbit calibration methodologies, approaches, and results derived from instrument on-board calibrators and lunar observations, as well as select Earth view targets. What is also discussed is the calibration consistency between MODIS and VIIRS and its potential impact on producing multi-sensor long-term data records. As illustrated, the overall performance of both MODIS and VIIRS continues to meet their design requirements.

MODIS

Monitoring VIIRS Thermal Emissive Bands Long-Term Performance Using Lunar Observations

Two Visible Infrared Imaging Radiometer Suite (VIIRS) instruments are currently operating in space, one onboard the Suomi National Polar-orbiting Partnership (S-NPP) satellite launched on October 28, 2011 and the other onboard the NOAA-20 satellite launched on November 18, 2017. The performance of the seven VIIRS thermal emissive bands (TEBs) is monitored via on-orbit calibration data, the analysis of Earth view observations, as well as inter-comparisons with other sensors. The Moon has been used as a unique invariant target for on-orbit sensor calibration because its surface property is considered extremely stable over time in spectra, radiometry, and geometry with a repeatable lunar phase. This paper presents the assessments of the S-NPP and NOAA-20 VIIRS TEB on-orbit calibration stability using near-monthly scheduled lunar observations over their respective missions. The methodology previously developed for MODIS is now applied to VIIRS to monitor the TEB long-term stability by extracting and trending the lunar brightness temperatures from selected unsaturated pixels. It includes a correction applied to reduce any impact from the Sun–Moon geometry (e.g. Sun-Moon distance variation). The results show that the mission-long brightness temperature trends from the lunar surface are stable for all VIIRS TEBs, despite noticeable seasonal variations in the lunar thermal emissive radiance.

VIIRS

MODIS TEB Calibration Algorithm Improvements for Future L1B Collection

Terra and Aqua MODIS have provided continuous global observations for scientific applications for more than 20 and 18 years, respectively. The radiometric calibration of the MODIS thermal emissive bands (TEB) is based on a quadratic approximation of the instrument response. The calibration coefficients look-up tables (LUTs) are updated using the response of the MODIS on-board blackbody (BB)with the response background subtracted by space view. The quarterly on-board BB warm-up and cool-down activity temperature ranges from 270 K to 315 K, and the derived offset has a relatively large uncertainty. Electronic cross-talk, an artifact that affects both instrument calibration and Earth view (EV) radiance retrievals, is corrected based on lunar observations. Calibration assessments using EV observations (e.g. ocean, desert, Antarctic Dome Concordia, and deep convective clouds) provide useful information to evaluate the impact of the Terra safe mode (February 2016) and Aqua MODIS formatter reset (January 2018) events on both MODIS instruments. This study focuses on the TEB radiometric calibration algorithm improvements for future collections based on calibration assessments using EV measurements and analytical modeling. Measurement stability and consistency over specific Earth scenes with a wide temperature range, as well as their brightness temperature (BT) dependency, are used for bias estimations in the calibration coefficients. Calibration coefficients are derived and updated after adjusting the current fitting algorithm. Thereafter, using the test LUTs, their impact on the Level 1B (L1B) data for different EV scenes is analyzed.

MODIS

The MODIS TEBs Calibration and Look-up Table Delivery Process for Collections 6 and 6.1

MODIS is a cross-track, whisk-broom scanning imaging radiometer with a double-sided scan mirror that collects data in 36 spectral bands. Sixteen of the 36 MODIS spectral bands are Thermal Emissive Bands (TEBs) whose spectral wavelengths range from 3.5 μm to 14.4 μm. The TEB detectors are calibrated on a scan-by-scan basis using a quadratic calibration algorithm by observing both the MODIS on-board blackbody (BB) and a background space view reference. Blackbody warm-up/cool-down (WUCD) events are performed quarterly to track on-orbit changes associated with the TEB detectors' non-linearity. Following each WUCD, the calibration coefficients in the quadratic algorithm, and their associated contributions to the total uncertainty, are updated and delivered through separate look-up tables (LUTs) when all update criteria are met. Afterwards, the LUTs are incorporated into the Level 1B (L1B) product. Since the Terra MODIS mission began, a steady increase in electronic cross-talk has been observed for TEBs 27{30. Starting from Collection 6.1, an algorithm has been applied using correction coefficients derived from regularly-scheduled lunar observations, with the correction LUT update dependent on its impact on the current L1B product. The MODIS Characterization Support Team (MCST) has established a comprehensive set of procedures to assure timely and accurate LUT updates, and maintain the quality and accuracy of the L1B and science products. This paper provides an overview of the current calibration and LUT delivery process for the MODIS TEBs in Collections 6 and 6.1.

MODIS

Comparison of the MODIS and VIIRS Thermal Emissive Band Radiometric Calibration

Moderate Resolution Imaging Spectroradiometer (MODIS) and Visible Infrared Imaging Radiometer Suite (VIIRS) are major instruments for Earth science observations.Nearly 40 MODIS scientific products and a wide range of VIIRS environmental data records are produced using their global observations. The consistency of the MODIS and VIIRS calibrated data is important for the study of Earth science. This article assesses the calibration consistency of the Aqua MODIS and VIIRS thermal emissive band (TEB) data. To remove the impact of the mismatched relative spectral response (RSR) on the comparisons, the simultaneous nadir observation data from the cross-track infrared sounder (CrIS) and the infrared atmospheric sounding interferometer (IASI) are used as references in two different methods to independently verify the consistency. The comparisons of the MODIS and VIIRS TEB calibrated data show that the brightness temperature (BT) differences for comparable bands between MODIS and VIIRS are in general within 0.2 K for BT larger than 230 K. The differences of all comparable MODIS and VIIRS TEBs are consistent over time. MODIS measurements agree better with N20 VIIRS than with Suomi National Polar-orbiting Partnership (S-NPP) VIIRS. The S-NPP VIIRS measurements are higher than N20 VIIRS, within 0.1 Kfor long-wave infrared (LWIR) bands and 0.3 K for band M13.

Calibration

Improvements in the on-orbit response versus scan-angle characterization for the MODIS ocean color bands

On-orbit characterization of the response versus scan-angle (RVS) is one of the most challenging aspects of the reflective solar band (RSB) calibration for the MODIS instruments onboard the Terra and Aqua spacecraft. The degradation of the solar diffuser, together with the lack of onboard calibrators (OBCs) to cover additional scan angles, other than the one for the lunar observations, has resulted in the use of Earth view responses from the pseudo-invariant desert sites to track the on-orbit RVS changes. This approach has been implemented in Collection 6 (C6) and C6.1 for bands 1-4, 8 and 9 of both instruments and also band 10 of Terra MODIS. As the missions continue to operate over a decade beyond their designed life and the instrument optics continue to degrade, it is expected that the OBC based RVS currently applied to other bands, specifically the high-gain ocean bands, will be inadequate to maintain the long-term calibration stability. An interband calibration approach is formulated and implemented in this paper. The proposed approach relies on the use of a spectrally matching stable reference band to evaluate the long-term calibration stability of the high-gain ocean bands that typically saturate while viewing the selected calibration deserts. Results from this approach indicate a noticeable drift for Terra MODIS bands 11 and 12 whereas the Aqua bands continue to show excellent temporal stability. These results are consistent with the corrections derived by the NASA Ocean Biology Processing Group (OBPG) and are expected to have minimum impacts on the downstream science products.

MODIS

Early Calibration and Performance Assessments of NOAA-20 VIIRS Thermal Emissive Bands

The Visible Infrared Imaging Radiometer Suite (VIIRS) sensor aboard the NOAA-20 (previously JPSS-1) spacecraft has successfully operated since its launch in November, 2017. Similar to the first VIIRS instrument on the Suomi-NPP spacecraft, data is collected in 22 spectral bands that are calibrated by a set of onboard calibrators. This paper provides an overview of the NOAA-20 VIIRS on-orbit operation and calibration, with a particular focus on the thermal emissive bands (TEBs). Results presented in this paper include the on-orbit changes in the TEB spectral band responses, detector noise characterization, and key calibration parameters, such as the non-linear coefficients derived from the blackbody warm-up cool-down cycle. Other issues, such as the early mission long-wave infrared (LWIR) response degradation due to icing on the dewar window, and their impact on sensor calibration are also discussed. Since launch, the VIIRS instrument temperature has been stable to within ± 0.8K and the cold focal plane temperatures are well controlled with variations less than 40 mK. With the exception of the early degradation observed in the LWIR bands, the TEB gains have been stable to within 0.04% (except I5 at 0.07%). Based on the current performance, VIIRS is expected to meet its calibration requirements throughout its design lifetime.

NEdT

Early Calibration and Performance Assessments of NOAA-20 VIIRS Thermal Emissive Bands

The Visible Infrared Imaging Radiometer Suite (VIIRS) sensor aboard the NOAA-20 (previously JPSS-1) spacecraft has successfully operated since its launch in November, 2017. Similar to the rst VIIRS instrument on the Suomi-National Polar-orbiting Partnership (SNPP) spacecraft, the data are collected in 22 spectral bands that are calibrated by a set of onboard calibrators. This paper provides an overview of the NOAA-20 VIIRS on-orbit operation and calibration, with a particular focus on the thermal emissive bands (TEBs). The results presented in this paper include the on-orbit changes in the TEB spectral band responses, detector noise characterization, and key calibration parameters, such as the nonlinear coefcients derived from the blackbody warm-up cool-down cycles. Other issues, such as the early mission long-wave infrared (LWIR) response degradation due to icing on the dewar window, and their impact on sensor calibration are also discussed. Since launch, the VIIRS instrument temperature has been stable to within ±0.8 K and the cold focal plane temperatures are well controlled with variations less than 40 mK. With the exception of the early degradation observed in the LWIR bands, the TEB gains have been stable to within 0.04% (except I5 at 0.07%). Based on the current performance, VIIRS is expected to meet its calibration requirements throughout its design lifetime.

noise equivalent 25 differential temperature (NEdT