Search NASA⌕ Search

NASA NTRS · 19720003678

A thermosphere composition measurement using a quadrupole mass spectrometer with a side energy focussing quasi-open ion source

Abstract

The atomic oxygen concentration in the altitude range 130 to 240 km was determined through the use of a quadrupole spectrometer with a strongly focussing ion source. The instrument is used in the Thermosphere Probe in a manner that greatly increases the proportion of measured oxygen ions that have not experienced a surface collision and permits quantitative evaluation of surface recombination and thermalization effects which inevitably enter all spectrometer determinations. The data obtained strengthen the concept that consideration of surface effects is significant in quantifying spectrometer measurements of reactive gases, and tend to be in agreement with von Zahn's recent results.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Niemann, H. B., Spencer, N. W., Schmitt, G. A.. 1971-07-01. A thermosphere composition measurement using a quadrupole mass spectrometer with a side energy focussing quasi-open ion source. https://ntrs.nasa.gov/citations/19720003678

Cite the original work for its findings. Save a collection to share your selection of sources.