Search NASA⌕ Search

NASA NTRS · 19720005703

Liquid-filled transient pressure measuring systems: A method for determining frequency response

Abstract

An equation is given and experimentally verified for computing the resonant frequency of liquid-filled transient pressure measuring systems. Resonant frequencies of 100 to 1000 Hz are typical of those systems tested. The effect of noncondensable gas bubbles on system response is described. A method for determining transducer volumetric compliance is presented. An example system is described and analyzed to demonstrate the use of the theory.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Anderson, R. C., Englund, D. R., Jr.. 1971-12-01. Liquid-filled transient pressure measuring systems: A method for determining frequency response. https://ntrs.nasa.gov/citations/19720005703

Cite the original work for its findings. Save a collection to share your selection of sources.