NASA NTRS · 19720010775
Wind tunnel mass spectrometer design study
Abstract
The sampling probe to measure the free stream gas composition in a hypersonic wind tunnel is described. Using a free expansion of the continuum flow, a series of sampling inlet-pumping stages is evaluated. Mass analysis is carried out with a quadrupole mass spectrometer employing a crossed electron beam, molecular beam ion source, and an off-axis electron multiplier. The spatial distribution of the sampled molecular flux and the pumping requirements are evaluated for varying inlet conditions of gas density and flow velocity. The signal-to-noise ratio of the molecular beam formed by this series of coaxial inlets is also determined. The signal-to-noise ratio of the unscattered flux is sufficiently high that an unambiguous determination of the beam composition and of the free stream gas properties can be made. The magnitude of the detected signal is sufficiently high that measurement of trace constituents two orders of magnitude below the primary beam signal can be made.
Keep this discovery
Explore connections, maps & timelines
Hueser, J. E., Fowler, P.. 1971-01-01. Wind tunnel mass spectrometer design study. https://ntrs.nasa.gov/citations/19720010775
Cite the original work for its findings. Save a collection to share your selection of sources.