Search NASA⌕ Search

NASA NTRS · 19720011866

Attitude-referenced radiometer study. Part 2: Primary calibration system

Abstract

A primary calibration system, PCS, for infrared radiometers has been developed, built, and tested. The system allows radiometers to be calibrated with less than 1 percent error for use in earth coverage horizon measurements, earth resources surveys, and synoptic meteorological measurement. The final design, fabrication and test of the PCS are reported. A detailed description of the PCS construction is presented, along with the results of a complete series of functional tests. Test to verify the source thermal characteristics, collimator reflectance, and output beam characteristics are described and their results presented.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Williamson, W. R., Otte, A. A.. 1971-12-01. Attitude-referenced radiometer study. Part 2: Primary calibration system. https://ntrs.nasa.gov/citations/19720011866

Cite the original work for its findings. Save a collection to share your selection of sources.