Search NASAโŒ• Search

NASA NTRS ยท 19720050973

The geometric factor of a cylindrical plate electrostatic analyzer.

Abstract

A method for calculating the geometric factor of cylindrical plate electrostatic energy analyzers with various detector geometries is described. The effects of the fringe field are estimated. For a special simple case an exact geometric factor is calculated, enabling an estimate to be made of the inaccuracies of the approximations used in other cases. The results of some calculations are presented, and a simple approximate expression for the geometric factor is deduced.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Johnstone, A. D.. 1972-07-01. The geometric factor of a cylindrical plate electrostatic analyzer.. https://ntrs.nasa.gov/citations/19720050973