Search NASAโŒ• Search

NASA NTRS ยท 19730010712

Calibration plan for ultraviolet spectrometer experiment S169

Abstract

The calibration test equipment presented is employed to calibrate completely assembled and flight-ready Apollo Ultraviolet Spectrometers (UVS) consisting of a vacuum system in which the UVS can be installed. In it a premonochromator produces a monochromatic beam at any desired wavelength from a UV source. This beam can be focused by an adjustable transfer mirror to any point on the entrance slit of the UVS. The transfer mirror can also direct the beam to one or both of two absolutely calibrated reference photomultiplier tubes to determine the number of photons which enter the UVS. A movable mirror rhomb is also provided to intermittently monitor the number of photons in the beam when the beam is centered on the entrance slit.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

1971-09-22. Calibration plan for ultraviolet spectrometer experiment S169. https://ntrs.nasa.gov/citations/19730010712