NASA NTRS ยท 19730036820
Vacuum system for an infrared calibration facility.
Abstract
Calibration facilities for large aperture (up to 56 cm diam) infrared radiometers require commeasurable size vacuum chambers with full thermal shrouding and ultraclean pumping systems. Such a facility has been fabricated and is described in detail. The chamber is 106 cm in diameter and 4.87 m long. It utilizes integral ion pumping for operation in the 1 microtorr region. Economic considerations and time constraints of the total system program required local fabrication and, therefore, nonproprietary-type flange sealing.The sealing mechanism used was the flat flange with aluminum wire effecting the seal. Included in the discussion is the description of the cryogen distribution system which allows complete constant temperature control of the several shroud structures.
Keep this discovery
Explore connections, maps & timelines
Uhlin, J.. 1973-02-01. Vacuum system for an infrared calibration facility.. https://ntrs.nasa.gov/citations/19730036820