NASA NTRS · 19730043833
The retarding-potential analyzer on Atmosphere Explorer.
Abstract
A planar retarding-potential analyzer will be included in each of the Atmosphere Explorer (AE) payloads. The primary functions of the instrument are to supply ion-temperature and ion-concentration data, which it will do at least once every 40 km of flight path in the region of interest with an expected accuracy of better than 2%. In addition, the instrument will determine ion-drift velocities and energy spectra of both the thermal and suprathermal electrons. During eclipse, negative-ion concentrations greater than 0.3/cu cm can be detected, though the normal positive-ion sensitivity is approximately 1.5 ions per cu cm. Changes in the ion concentration along the flight path greater than 0.1% can be monitored with a spatial resolution of less than 40 m.
Keep this discovery
Explore connections, maps & timelines
Hanson, W. B., Zuccaro, D. R., Lippincott, C. R., Sanatani , S.. 1973-04-01. The retarding-potential analyzer on Atmosphere Explorer.. https://ntrs.nasa.gov/citations/19730043833
Cite the original work for its findings. Save a collection to share your selection of sources.