Abstract

Results obtained from several basic measurements of the performance of an SEC vidicon camera head are presented. The measurements or tests are patterned around the sensor requirements which exist for a rapid response coronagraph. Following a brief description of the SEC vidicon camera head, test set-up, and operating procedures, results are presented for five basic tests. These tests include: (1) Input-output, or transfer, characteristics; (2) modulation transfer function; (3) signal-to-noise ratio; (4) photocathode sensitivity; and (5) target gain. The input-output characteristics and photocathode sensitivity measurements were obtained for three bandpasses centered about 4000, 5250, and 6200 A. Results are also given for the variation of photocathode sensitivity with position on the cathode surface.