Abstract
A low frost-point humidity generator has been developed at NBS to provide a capability for calibration, testing, and research at very low levels of water vapor content in such gases as atmospheric air, carbon dioxide and nitrogen. The generator produces frost points from -30 to -100 C at ambient pressures from 500 to 200,000 pascals (0.005 to 2 atm.). This is equivalent to mixing ratios of 4 micrograms to 51 grams of water vapor per kilogram of dry air and to vapor pressures of .0014 to 38 pascals. The generated test gas can be fed to a test chamber with independent temperature control between +25 and -100 C. The uncertainty of the frost point in the test chamber is estimated not to exceed 0.05 deg C.
Keep this discovery
Explore connections, maps & timelines
Greenspan, L.. 1973-10-01. Low frost-point humidity generator. https://ntrs.nasa.gov/citations/19740041906