Search NASAโŒ• Search

NASA NTRS ยท 19740050682

Preliminary results with saturable microchannel array plates

Abstract

Microchannel array plates with a performance comparable to that of a conventional channel electron multiplier have been obtained for the first time. These array plates employ an angled electrostatic field to inhibit the feedback of positive ions within the microchannels. Saturated output pulse height distributions with modal gain values in excess of 10 million have been obtained and stable operation demonstrated over a range of ambient pressures from 0.0000001 to 0.00008 torr. However, a time-dependent reduction in the gain has been observed with these experimental plates because of the accumulation of charge on the insulating strips which are inserted in the wall of the microchannel to establish the angled electrostatic field.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Timothy, J. G.. 1974-06-01. Preliminary results with saturable microchannel array plates. https://ntrs.nasa.gov/citations/19740050682