The hazards inherent in the Portable Medical Status System are identified, and the measures taken to reduce them to an acceptable level are described. Identification of these hazards is a prerequisite to use of the system on humans in the earth environment. One hazard which is insufficiently controlled and which is considered a constraint to use on humans is the level of current possible in the electrodes for the EEG (electroencephalograph) circuitry. It exceeds the maximum specified. A number of procedural and design recommendations for enhancement of safety are made.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Lindsey, O. C.. 1976-02-01. Portable medical status system. https://ntrs.nasa.gov/citations/19760015805