Search NASA⌕ Search

NASA NTRS · 19760016007

Electrophoresis experiment, experiment MA-014

Abstract

A continuous, free flow electrophoresis study was conducted during the Apollo Soyuz Test Project mission to investigate and evaluate the increase in sample flow rate and sample resolution achievable in space. The electrophoresis equipment was designed for the separation of four mixtures of biological cells with variable sample flow rates, buffer flow rates, and electric field gradients. Separation quality was assessed by measuring the light from a quartz lamp through the electrophoresis channel and on to a photodiode system. The preliminary results indicate that all monitored systems operated correctly during the experiment. The optical system produced a light that was too bright to discern true cell distributions, but data were received that show a distribution of separated cells.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Hannig, K., Wirth, H.. 1976-02-01. Electrophoresis experiment, experiment MA-014. https://ntrs.nasa.gov/citations/19760016007

Cite the original work for its findings. Save a collection to share your selection of sources.