The purpose of this investigation was to develop two chromatographic columns which would enable a mass spectral identification of 40 specified compounds. The columns are for use in a toxic gas analyzer, which incorporates an automated gas chromatograph-mass spectrometer. Different types of stationary phases were investigated. The columns used were of the open tubular capillary type and were made of nickel. Limitations of initial and final temperature of operation led to final development of a column which could resolve most of the compounds required. The few unresolved components are capable of resolution and identification by the mass spectrometer. The columns (182m Ni x 0.8m 0.D x 0.5mm I.D) coated with Witconal La 23, yielded in excess of 200,000 theoretical plates and completed the analysis in less than 90 minutes using a carrier gas flow rate of 4 cc/min hydrogen.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Zlatkis, A.. 1976-09-01. Flight contaminant trace analyser. Phase 1: Chromatographic input system. https://ntrs.nasa.gov/citations/19760024218