Space Shuttle Trace Gas Analyzer

Abstract

A Trace Gas Analyzer (TGA) with the ability to detect the presence of toxic contaminants in the Space Shuttle atmosphere within the subparts-per-million range is under development. The design is a modification of the miniaturized Gas Chromatograph-Mass Spectrometer (GCMS) developed for the Viking Mars Lander. An ambient air sample is injected onto the GC column from a constant volume sample loop and separated into individual compounds for identification by the MS. The GC-MS interface consists of an effluent divider and a silver-paladium separator, an electrochemical cell which removes more than 99.99% of the hydrogen carrier gas. The hydrogen is reclaimed and repressurized without affecting the separator efficiency, a feature which enables a considerable weight reduction in the carrier gas supply system.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Dencker, W.. 1975-08-01. Space Shuttle Trace Gas Analyzer. https://searchnasa.com/record/nasa%3A19760029903/space-shuttle-trace-gas-analyzer