Search NASA⌕ Search

NASA NTRS · 19760039821

Linear polarimeter with rapid modulation, achromatic in the 0.3-1.1-micron range

Abstract

A polarimeter is described, which provides achromatic modulation over the spectral range 0.3 to 1.1 microns, with very small light losses. The instrument is the first one to use superachromatic half-wave plates; such a plate is made of three achromatic half-wave plates of quartz and magnesium fluoride cemented with the optical axis of the central plate making an angle of 1.00 rad with the axes of the parallel-oriented outer half-wave plates. Wavelength dependence of the axis direction of the optical axis was eliminated by using two identical such plates in front of a Wollaston prism. The plate closer to the prism is stationary, while the other is continuously rotated by a stepper motor. In the polarimeter constructed, the range of instrumentally induced wavelength dependence of the plane of polarization did not exceed 0.002 rad. Using the polarimeter on a 228-cm telescope, a 10-min integration through the blue filter of the UBV photometric system for a star of magnitude 10.0 gives the percentage polarization with a mean error of + or - 0.014%.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Frecker, J. E., Serkowski, K.. 1976-03-01. Linear polarimeter with rapid modulation, achromatic in the 0.3-1.1-micron range. https://ntrs.nasa.gov/citations/19760039821

Cite the original work for its findings. Save a collection to share your selection of sources.